An investigation on the power energy injection in a resonator by electron stream is reported in this paper, by documenting the power transfer between the electron current to a resonant system. The beam current emitted by an electron linear accelerator (LINAC) has been monitored through a particular radiation detector. This device was developed for the dose measurements of a Medical Linear LINAC where monitoring system are required for the real time control of the dose delivered to the patient
Leggieri, A., Passi, D., DI PAOLO, F., Felici, G., Ciccotelli, A., De Stefano, S., et al. (2015). Electron Beam Detection By Induced Resonance In Cylindrical Cavity. In PIERS 2015 Proceedings. Praga.
Electron Beam Detection By Induced Resonance In Cylindrical Cavity
DI PAOLO, FRANCO;
2015-08-01
Abstract
An investigation on the power energy injection in a resonator by electron stream is reported in this paper, by documenting the power transfer between the electron current to a resonant system. The beam current emitted by an electron linear accelerator (LINAC) has been monitored through a particular radiation detector. This device was developed for the dose measurements of a Medical Linear LINAC where monitoring system are required for the real time control of the dose delivered to the patientFile | Dimensione | Formato | |
---|---|---|---|
8 - FERMAT - Cavity.pdf
accesso aperto
Licenza:
Non specificato
Dimensione
1.71 MB
Formato
Adobe PDF
|
1.71 MB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.