Urinary tract infection is a common disease diagnosed from symptoms and clinical signs, and bacterial count per volume of urine. This study have evaluated the BiesseBioscreen analyzer as a new way to analyze urine samples enabling fast screening of urine, prior to reference standard methods currently utilized in microbiology analysis laboratory. We analyzed 962 urine samples from outpatients and inpatients of the Tor Vergata (TV) University Hospital of the University of Rome “Tor Vergata”. All samples were processed both with the BiesseBioscreen and with the standard methodology adopted by the clinical microbiology laboratory of TV Hospital and the results were compared. Of the samples analyzed 54.9% were concordant negative with the reference method and 21.6% concordant positive, 23.3% resulted false positive and 0.2% false negative. The results obtained from BiesseBioscreen showed a sensitivity of 99.0%, indicating it as a system suitable to rule out urinary tract infection. BiesseBioscreen could represent a valid method for screening negative samples to exclude from culture test with a potential reduction in time, workload and costs of the diagnosis.

Nicolai, E., Garau, S., Favalli, C., D'Agostini, C., Gratton, E., Motolese, G., et al. (2014). Evaluation of BiesseBioscreen as a new methodology for bacteriuria screening. NEW MICROBIOLOGICA, 37, 495-501.

Evaluation of BiesseBioscreen as a new methodology for bacteriuria screening

NICOLAI, ELEONORA;FAVALLI, CARTESIO;D'AGOSTINI, CARTESIO;ROSATO, NICOLA
2014-10-01

Abstract

Urinary tract infection is a common disease diagnosed from symptoms and clinical signs, and bacterial count per volume of urine. This study have evaluated the BiesseBioscreen analyzer as a new way to analyze urine samples enabling fast screening of urine, prior to reference standard methods currently utilized in microbiology analysis laboratory. We analyzed 962 urine samples from outpatients and inpatients of the Tor Vergata (TV) University Hospital of the University of Rome “Tor Vergata”. All samples were processed both with the BiesseBioscreen and with the standard methodology adopted by the clinical microbiology laboratory of TV Hospital and the results were compared. Of the samples analyzed 54.9% were concordant negative with the reference method and 21.6% concordant positive, 23.3% resulted false positive and 0.2% false negative. The results obtained from BiesseBioscreen showed a sensitivity of 99.0%, indicating it as a system suitable to rule out urinary tract infection. BiesseBioscreen could represent a valid method for screening negative samples to exclude from culture test with a potential reduction in time, workload and costs of the diagnosis.
1-ott-2014
Pubblicato
Rilevanza internazionale
Articolo
Comitato scientifico
Settore MED/07 - MICROBIOLOGIA E MICROBIOLOGIA CLINICA
English
Con Impact Factor ISI
KEY WORDS: Bacteriuria screening, Fluorescent assays, Urinary tract infection, Nucleic probe, Bacterial culture.
Nicolai, E., Garau, S., Favalli, C., D'Agostini, C., Gratton, E., Motolese, G., et al. (2014). Evaluation of BiesseBioscreen as a new methodology for bacteriuria screening. NEW MICROBIOLOGICA, 37, 495-501.
Nicolai, E; Garau, S; Favalli, C; D'Agostini, C; Gratton, E; Motolese, G; Rosato, N
Articolo su rivista
File in questo prodotto:
File Dimensione Formato  
Micro4_08_Nicolai.pdf

accesso aperto

Descrizione: pdf
Dimensione 73.87 kB
Formato Adobe PDF
73.87 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/102068
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 8
  • ???jsp.display-item.citation.isi??? 10
social impact