Abstract Introduction: This ex vivo study evaluated the accuracy of the Root ZX electronic apex locator (EAL) (J Morita Corp, Kyoto, Japan) in determining the working length during retreatment of canals sealed with 2 different carrier-based obturating materials (ProTaper Obturator [Dentsply Maillefer, Ballaigues, Switzerland] and GuttaCore [Dentsply Maillefer]) and also evaluated whether they influenced its accuracy differently. Methods: Fifty extracted single-rooted human teeth with sound apices were selected for the study. We measured canals with the direct visual technique using a #10 K-file and shaped them with ProTaper Universal instruments (Dentsply Maillefer) up to this predetermined length. After instrumentation, we determined the canal length again using the visual technique (direct length [DL]). This value was considered the ‘‘reference point.’’ Specimens were then placed in a tooth holder (Pro-Train; Simit Dental, Mantova, Italy), and the canal length was electronically measured (electronic length 1 [EL1]). Specimens were then obturated with ProTaper Obturators (group 1, n = 25) and GuttaCore (group 2, n = 25). Seven days later, a new electronic length was determined (electronic length 2 [EL2]) during retreatment procedures in the presence of the obturating material. EL1 and EL2 values were compared with DL using the Bland and Altman method. The different influences of the tested materials on the accuracy of the EAL were calculated with repeated measures analysis of variance. Results: Both EL1 and EL2 values provided a statistically significant overestimation of the actual canal length (DL [P < .05]), with the EL2 values higher from DL. The 2 different materials did not influence the accuracy of the EAL differently (P = .486). Conclusions: The measurements obtained with the EAL tested during orthograde retreatments can lead clinicians to overinstrumentation and consequent overfilling of the endodontic space. (J Endod 2014;40:2061–2065)

Mancini, M., Palopoli, P., Iorio, L., Conte, G., Cianconi, L. (2014). Accuracy of an electronic apex locator in the retreatment of teeth obturated with plastic or cross-linked gutta-percha carrier-based materials: An ex vivo study. JOURNAL OF ENDODONTICS, 40(12), 2061-2065.

Accuracy of an electronic apex locator in the retreatment of teeth obturated with plastic or cross-linked gutta-percha carrier-based materials: An ex vivo study.

CIANCONI, LUIGI
2014-12-01

Abstract

Abstract Introduction: This ex vivo study evaluated the accuracy of the Root ZX electronic apex locator (EAL) (J Morita Corp, Kyoto, Japan) in determining the working length during retreatment of canals sealed with 2 different carrier-based obturating materials (ProTaper Obturator [Dentsply Maillefer, Ballaigues, Switzerland] and GuttaCore [Dentsply Maillefer]) and also evaluated whether they influenced its accuracy differently. Methods: Fifty extracted single-rooted human teeth with sound apices were selected for the study. We measured canals with the direct visual technique using a #10 K-file and shaped them with ProTaper Universal instruments (Dentsply Maillefer) up to this predetermined length. After instrumentation, we determined the canal length again using the visual technique (direct length [DL]). This value was considered the ‘‘reference point.’’ Specimens were then placed in a tooth holder (Pro-Train; Simit Dental, Mantova, Italy), and the canal length was electronically measured (electronic length 1 [EL1]). Specimens were then obturated with ProTaper Obturators (group 1, n = 25) and GuttaCore (group 2, n = 25). Seven days later, a new electronic length was determined (electronic length 2 [EL2]) during retreatment procedures in the presence of the obturating material. EL1 and EL2 values were compared with DL using the Bland and Altman method. The different influences of the tested materials on the accuracy of the EAL were calculated with repeated measures analysis of variance. Results: Both EL1 and EL2 values provided a statistically significant overestimation of the actual canal length (DL [P < .05]), with the EL2 values higher from DL. The 2 different materials did not influence the accuracy of the EAL differently (P = .486). Conclusions: The measurements obtained with the EAL tested during orthograde retreatments can lead clinicians to overinstrumentation and consequent overfilling of the endodontic space. (J Endod 2014;40:2061–2065)
dic-2014
Pubblicato
Rilevanza internazionale
Articolo
Esperti anonimi
Settore MED/28 - MALATTIE ODONTOSTOMATOLOGICHE
English
Con Impact Factor ISI
Electronic apex locator; GuttaCore; ProTaper Obturator; retreatment; working length;
Mancini, M., Palopoli, P., Iorio, L., Conte, G., Cianconi, L. (2014). Accuracy of an electronic apex locator in the retreatment of teeth obturated with plastic or cross-linked gutta-percha carrier-based materials: An ex vivo study. JOURNAL OF ENDODONTICS, 40(12), 2061-2065.
Mancini, M; Palopoli, P; Iorio, L; Conte, G; Cianconi, L
Articolo su rivista
File in questo prodotto:
File Dimensione Formato  
JOE-40-12-2061.pdf

accesso aperto

Descrizione: Articolo Fulltext
Dimensione 309.69 kB
Formato Adobe PDF
309.69 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/99880
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 8
  • ???jsp.display-item.citation.isi??? 7
social impact