Background: This study aims to evaluate the repeatability of the Pentacam HR, comparing two different measurement modes (50-cornea fine and 25-3D scan) in patients affected by keratoconus. Methods: Multicenter retrospective study, conducted at Eye Clinic of the ASST-Spedali Civili-University of Brescia, Italy, and St. Paul’s Eye Unit, Royal Liverpool University Hospital, United Kingdom. A total of 72 eyes from 72 patients with keratoconus underwent six consecutive measurements, three using the 25-3D scan mode and three with the 50-Cornea fine mode. Measurements were made by one single observer, using the Scheimpflug corneal tomographer camera (Pentacam HR, Oculus, Wetzlar, Germany). Repeatability was assessed using the within-subject SD (Sw) statistic from a two-way analysis of variance. Results: Both measurement modes had excellent repeatability. The interclass coefficient correlation (ICC) was excellent (>0.9) in all the parameters evaluated, apart from anterior and posterior astigmatic axes and posterior astigmatism (ICC > 0.8) and index of height asymmetry (IHA) (ICC < 0.6). However, in 18 of 29 parameters, the ICC was higher in case of 25-3D scan. Repeatability limit for Kmax was 1.00D in 25-3D scan mode and 1.02D in case of 50-cornea fine. Conclusions: 25-3D scan may be preferable to 50-Cornea fine, in view of having slightly higher ICC in case of patients with keratoconus. Repeatability limits reported may be helpful in clinical practice for assessing the progression of keratoconus.

Romano, D., Coco, G., Borgia, A., Calza, ., S, K., S, ., et al. (2025). Repeatability of Pentacam HR in Keratoconus According to Two Different Scan Protocols: 25-3D Scan and 50-Cornea Fine. JOURNAL OF CLINICAL MEDICINE, 14(2) [10.3390/jcm14020439].

Repeatability of Pentacam HR in Keratoconus According to Two Different Scan Protocols: 25-3D Scan and 50-Cornea Fine

Coco, G;Romano, V
2025-01-11

Abstract

Background: This study aims to evaluate the repeatability of the Pentacam HR, comparing two different measurement modes (50-cornea fine and 25-3D scan) in patients affected by keratoconus. Methods: Multicenter retrospective study, conducted at Eye Clinic of the ASST-Spedali Civili-University of Brescia, Italy, and St. Paul’s Eye Unit, Royal Liverpool University Hospital, United Kingdom. A total of 72 eyes from 72 patients with keratoconus underwent six consecutive measurements, three using the 25-3D scan mode and three with the 50-Cornea fine mode. Measurements were made by one single observer, using the Scheimpflug corneal tomographer camera (Pentacam HR, Oculus, Wetzlar, Germany). Repeatability was assessed using the within-subject SD (Sw) statistic from a two-way analysis of variance. Results: Both measurement modes had excellent repeatability. The interclass coefficient correlation (ICC) was excellent (>0.9) in all the parameters evaluated, apart from anterior and posterior astigmatic axes and posterior astigmatism (ICC > 0.8) and index of height asymmetry (IHA) (ICC < 0.6). However, in 18 of 29 parameters, the ICC was higher in case of 25-3D scan. Repeatability limit for Kmax was 1.00D in 25-3D scan mode and 1.02D in case of 50-cornea fine. Conclusions: 25-3D scan may be preferable to 50-Cornea fine, in view of having slightly higher ICC in case of patients with keratoconus. Repeatability limits reported may be helpful in clinical practice for assessing the progression of keratoconus.
11-gen-2025
Pubblicato
Rilevanza internazionale
Articolo
Esperti anonimi
Settore MEDS-17/A - Malattie dell'apparato visivo
English
Pentacam; cornea; keratoconus; precision; repeatability; tomography; topography
Romano, D., Coco, G., Borgia, A., Calza, ., S, K., S, ., et al. (2025). Repeatability of Pentacam HR in Keratoconus According to Two Different Scan Protocols: 25-3D Scan and 50-Cornea Fine. JOURNAL OF CLINICAL MEDICINE, 14(2) [10.3390/jcm14020439].
Romano, D; Coco, G; Borgia, A; Calza, ; S, K; S, ; Gadhvi, K; Semeraro, F; Romano, V
Articolo su rivista
File in questo prodotto:
File Dimensione Formato  
2025. Repeatability of Pentacam HR in Keratoconus According to Two Different Scan Protocols- 5 25-3D Scan and 50-Cornea Fine.pdf

accesso aperto

Tipologia: Versione Editoriale (PDF)
Licenza: Creative commons
Dimensione 2.13 MB
Formato Adobe PDF
2.13 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/458985
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 1
social impact