Background: International guidelines recommend the use of high-sensitivity cardiac troponin (hs-cTn) I and T methods for the detection of myocardial injury as a pre-requisite for the diagnosis of acute myocardial infarction (AMI) in patients admitted to the emergency department. Recently, Mindray (Mindray Bio-Medical Electronics Co., Ltd., Shenzhen, China) has introduced a new chemiluminescence immunoassay (CLIA) for the detection of the cTn complex. The present study aims to verify and validate the hs-cTnI Mindray assay on the new automated CL2600i analyzer compared to the routine Alinity-i series instrument by Abbott (Abbott, Chicago, IL, USA). Methods: This study evaluated linearity, precision through the 5 x 5 protocol, methodological comparison on plasma and serum matrices, hs-cTnI 99th percentile imprecision, and the hs-cTnI detection rate in a healthy population. Results: The results obtained proved that the performance of the Mindray hs-cTnI test on the CL2600i platform was closely comparable to the Abbott Alinity-i system (plasma R2: 0.974; serum R2: 0.995). The CVs were consistently low, and no significant differences were reported. Excellent analytical performance, with high sensitivity, was also observed in the healthy population (overall detection rate: 79%), as well as good linearity within the measuring range (R2: 0.994). Conclusions: The Mindray hs-cTnI test confirms its robustness and utility in routine practice as an advanced assay. The new technology, with more sensitive detection methods, may improve the accuracy and reliability of cardiac biomarker testing, ultimately leading to better outcomes in the management of patients with AMI and other cardiac conditions.
Tomassetti, F., Fiorelli, D., Cappa, E., Giovannelli, A., Pelagalli, M., Velocci, S., et al. (2025). Performance evaluation of the new high-sensitivity cardiac troponin I immunoassay on CL-2600i Mindray analyzer. DIAGNOSTICS, 15(8) [10.3390/diagnostics15081031].
Performance evaluation of the new high-sensitivity cardiac troponin I immunoassay on CL-2600i Mindray analyzer
Tomassetti F.;Fiorelli D.;Cappa E.;Giovannelli A.;Minieri M.;Perrone M. A.;
2025-01-01
Abstract
Background: International guidelines recommend the use of high-sensitivity cardiac troponin (hs-cTn) I and T methods for the detection of myocardial injury as a pre-requisite for the diagnosis of acute myocardial infarction (AMI) in patients admitted to the emergency department. Recently, Mindray (Mindray Bio-Medical Electronics Co., Ltd., Shenzhen, China) has introduced a new chemiluminescence immunoassay (CLIA) for the detection of the cTn complex. The present study aims to verify and validate the hs-cTnI Mindray assay on the new automated CL2600i analyzer compared to the routine Alinity-i series instrument by Abbott (Abbott, Chicago, IL, USA). Methods: This study evaluated linearity, precision through the 5 x 5 protocol, methodological comparison on plasma and serum matrices, hs-cTnI 99th percentile imprecision, and the hs-cTnI detection rate in a healthy population. Results: The results obtained proved that the performance of the Mindray hs-cTnI test on the CL2600i platform was closely comparable to the Abbott Alinity-i system (plasma R2: 0.974; serum R2: 0.995). The CVs were consistently low, and no significant differences were reported. Excellent analytical performance, with high sensitivity, was also observed in the healthy population (overall detection rate: 79%), as well as good linearity within the measuring range (R2: 0.994). Conclusions: The Mindray hs-cTnI test confirms its robustness and utility in routine practice as an advanced assay. The new technology, with more sensitive detection methods, may improve the accuracy and reliability of cardiac biomarker testing, ultimately leading to better outcomes in the management of patients with AMI and other cardiac conditions.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.


