Background: the Cocoon patent foramen ovale (PFO) Occluder is a new generation nitinol alloy double-disk device coated with nanoplatinum, likely useful in patients with nickel hypersensitivity. Early results and mid-term outcomes of this device in percutaneous PFO closure are missing. Aims: to assess the preliminary ecacy and safety profile of PFO closure with Cocoon device in an Italian multi-center registry. Methods: This is a prospective registry of 189 consecutive adult patients treated with the Cocoon PFO Occluder at 15 Italian centers from May 2017 till May 2020. Patients were followed up for 2 years. Results: closure of the PFO with Cocoon Occluder was carried out successfully in all patients, with complete closure without residual shunt in 94.7% of the patients and minimal shunt in 5.3%. Except from a case of paroxysmal supraventricular tachycardia and a major vascular bleeding, no procedural and in-hospital device-related complications occurred. No patient developed cardiac erosions, allergic reactions to nickel, or any other major complications during the follow-up. During the follow-up period, 2 cases of new-onset atrial fibrillation occurred within thirty-d
Testa, L., Popolo Rubbio, A., Squillace, M., Albano, F., Cesario, V., Casenghi, M., et al. (2023). Patent foramen ovale occlusion with the Cocoon PFO Occluder. The PROS-IT collaborative project. FRONTIERS IN CARDIOVASCULAR MEDICINE, 9 [10.3389/fcvm.2022.1064026].
Patent foramen ovale occlusion with the Cocoon PFO Occluder. The PROS-IT collaborative project
Ussia G. P.;Sangiorgi G.;
2023-01-01
Abstract
Background: the Cocoon patent foramen ovale (PFO) Occluder is a new generation nitinol alloy double-disk device coated with nanoplatinum, likely useful in patients with nickel hypersensitivity. Early results and mid-term outcomes of this device in percutaneous PFO closure are missing. Aims: to assess the preliminary ecacy and safety profile of PFO closure with Cocoon device in an Italian multi-center registry. Methods: This is a prospective registry of 189 consecutive adult patients treated with the Cocoon PFO Occluder at 15 Italian centers from May 2017 till May 2020. Patients were followed up for 2 years. Results: closure of the PFO with Cocoon Occluder was carried out successfully in all patients, with complete closure without residual shunt in 94.7% of the patients and minimal shunt in 5.3%. Except from a case of paroxysmal supraventricular tachycardia and a major vascular bleeding, no procedural and in-hospital device-related complications occurred. No patient developed cardiac erosions, allergic reactions to nickel, or any other major complications during the follow-up. During the follow-up period, 2 cases of new-onset atrial fibrillation occurred within thirty-d| File | Dimensione | Formato | |
|---|---|---|---|
|
Patent foramen ovale.pdf
accesso aperto
Tipologia:
Versione Editoriale (PDF)
Licenza:
Creative commons
Dimensione
435.89 kB
Formato
Adobe PDF
|
435.89 kB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.