Purpose Planar cardiac scintigraphy with [123I]I-mIBG is widely used to image myocardial sympathetic activity. In multicentre studies, different γ-cameras can cause variability in quantitative parameters. This study aimed to harmonise multicentre [123I]I-mIBG data using a custom-designed phantom and to assess its feasibility and acceptability. Methods A ‘tube and bottle’ phantom was designed to standardise the heart-to-mediastinum (H/M) ratio across 15 centres. Each centre prepared three versions of the phantom (A, B, and C) with varying pertechnetate ([99mTc]Tc-O4-) activities, acquired static images using their own γ-camera, and uploaded DICOM data to a shared platform. In the phantom, the tube and bottle represent the heart and mediastinum, respectively, with the tube-to-bottle (T/B) ratio simulating the H/M ratio. The reference centre analysed the images and calculated the T/B ratios, applying linear regression for data harmonisation. A survey was conducted to assess the phantom’s usability. Results The harmonised T/B ratios for versions A and B were 20.46 ± 0.78 and 6.19 ± 0.39, respectively. The average slope and intercept of the regression line across the participating centres resulted in 1.07 ± 0.38 and − 0.82 ± 4.95. Survey feedback indicated high feasibility and acceptability, with all centres recommending the phantom for multicentre harmonisation. Conclusions The custom-made phantom effectively harmonised experimental data across different centres, supporting its use in multicentre studies to improve data consistency.
Filippi, L., Bianconi, F., Minestrini, M., De Vincentis, G., Frantellizzi, V., Marongiu, A., et al. (2025). Multi-centre data harmonisation applied to heart-to-mediastinum quantification in parkinsonism (ITA-MIBG): a cross-calibration phantom study with tube and bottle. CLINICAL AND TRANSLATIONAL IMAGING, 1-8 [10.1007/s40336-025-00681-4].
Multi-centre data harmonisation applied to heart-to-mediastinum quantification in parkinsonism (ITA-MIBG): a cross-calibration phantom study with tube and bottle
Filippi, Luca;Schillaci, Orazio;
2025-02-04
Abstract
Purpose Planar cardiac scintigraphy with [123I]I-mIBG is widely used to image myocardial sympathetic activity. In multicentre studies, different γ-cameras can cause variability in quantitative parameters. This study aimed to harmonise multicentre [123I]I-mIBG data using a custom-designed phantom and to assess its feasibility and acceptability. Methods A ‘tube and bottle’ phantom was designed to standardise the heart-to-mediastinum (H/M) ratio across 15 centres. Each centre prepared three versions of the phantom (A, B, and C) with varying pertechnetate ([99mTc]Tc-O4-) activities, acquired static images using their own γ-camera, and uploaded DICOM data to a shared platform. In the phantom, the tube and bottle represent the heart and mediastinum, respectively, with the tube-to-bottle (T/B) ratio simulating the H/M ratio. The reference centre analysed the images and calculated the T/B ratios, applying linear regression for data harmonisation. A survey was conducted to assess the phantom’s usability. Results The harmonised T/B ratios for versions A and B were 20.46 ± 0.78 and 6.19 ± 0.39, respectively. The average slope and intercept of the regression line across the participating centres resulted in 1.07 ± 0.38 and − 0.82 ± 4.95. Survey feedback indicated high feasibility and acceptability, with all centres recommending the phantom for multicentre harmonisation. Conclusions The custom-made phantom effectively harmonised experimental data across different centres, supporting its use in multicentre studies to improve data consistency.File | Dimensione | Formato | |
---|---|---|---|
s40336-025-00681-4.pdf
accesso aperto
Tipologia:
Versione Editoriale (PDF)
Licenza:
Copyright dell'editore
Dimensione
1.08 MB
Formato
Adobe PDF
|
1.08 MB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.