Background: Conventional stented, rapid deployment and new-generation stented valves are now available for surgical aortic valve replacement (SAVR). New-generation devices feature advanced tissue treatment for theoretical prolonged durability and a new stent design able to expand in case of future transcatheter Valve-in-Valve. Aim of this retrospective, multicenter, propensity-weighted study was to compare early clinical and hemodynamic outcomes of these three different bioprostheses. Methods: We analyzed data of 2589 patients from two national multicenter registries and one Institutional database. Study devices were Magna Ease, Intuity/Intuity Elite and Inspiris Resilia (Edwards Lifesciences, Irvine, CA, USA) and were implanted in 296 (11.4 %), 1688 (65.2 %) and 605 (23.4 %) patients, respectively. A propensity score weighting approach was employed. Results: In isolated SAVR, aortic cross clamp (ACC) time was shorter for Intuity (Magna Ease: 87, Intuity: 55, Inspiris: 70 min; Magna Ease vs. Intuity: p < 0.001; Inspiris vs. Intuity: p < 0.001). Overall mortality was 2 %, 1.7 % and 0.5 % in Magna Ease, Intuity and Inspiris groups, respectively (Magna Ease vs. Intuity: p = 0.476; Inspiris vs. Intuity: p = 0.395); permanent pace-maker implantation rate was lower for Inspiris (Magna Ease: 6 %, Intuity: 6 %, Inspiris: 2 %; Magna Ease vs. Intuity: p = 0.679; Inspiris vs. Intuity: p < 0.001). Median mean gradients were 13, 10 and 10 mmHg for Magna Ease, Intuity and Inspiris, respectively (Magna Ease vs. Intuity: p < 0.001; Inspiris vs. Intuity: p = 0.13). Conclusions: All study devices provide excellent early clinical and hemodynamic outcomes. Inspiris shows low rates of permanent pace-maker implantation and its transaortic gradients are similar to rapid-deployment valves and lower than Magna Ease.

D'Onofrio, A., Cibin, G., Tessari, C., Lorenzoni, G., Luzi, G., Manzan, E., et al. (2024). Multicenter, propensity-weighted comparison of stented, rapid-deployment and new-generation aortic valves. INTERNATIONAL JOURNAL OF CARDIOLOGY. HEART & VASCULATURE, 54 [10.1016/j.ijcha.2024.101487].

Multicenter, propensity-weighted comparison of stented, rapid-deployment and new-generation aortic valves

D'Onofrio A.
Writing – Original Draft Preparation
;
2024-01-01

Abstract

Background: Conventional stented, rapid deployment and new-generation stented valves are now available for surgical aortic valve replacement (SAVR). New-generation devices feature advanced tissue treatment for theoretical prolonged durability and a new stent design able to expand in case of future transcatheter Valve-in-Valve. Aim of this retrospective, multicenter, propensity-weighted study was to compare early clinical and hemodynamic outcomes of these three different bioprostheses. Methods: We analyzed data of 2589 patients from two national multicenter registries and one Institutional database. Study devices were Magna Ease, Intuity/Intuity Elite and Inspiris Resilia (Edwards Lifesciences, Irvine, CA, USA) and were implanted in 296 (11.4 %), 1688 (65.2 %) and 605 (23.4 %) patients, respectively. A propensity score weighting approach was employed. Results: In isolated SAVR, aortic cross clamp (ACC) time was shorter for Intuity (Magna Ease: 87, Intuity: 55, Inspiris: 70 min; Magna Ease vs. Intuity: p < 0.001; Inspiris vs. Intuity: p < 0.001). Overall mortality was 2 %, 1.7 % and 0.5 % in Magna Ease, Intuity and Inspiris groups, respectively (Magna Ease vs. Intuity: p = 0.476; Inspiris vs. Intuity: p = 0.395); permanent pace-maker implantation rate was lower for Inspiris (Magna Ease: 6 %, Intuity: 6 %, Inspiris: 2 %; Magna Ease vs. Intuity: p = 0.679; Inspiris vs. Intuity: p < 0.001). Median mean gradients were 13, 10 and 10 mmHg for Magna Ease, Intuity and Inspiris, respectively (Magna Ease vs. Intuity: p < 0.001; Inspiris vs. Intuity: p = 0.13). Conclusions: All study devices provide excellent early clinical and hemodynamic outcomes. Inspiris shows low rates of permanent pace-maker implantation and its transaortic gradients are similar to rapid-deployment valves and lower than Magna Ease.
2024
Pubblicato
Rilevanza internazionale
Articolo
Comitato scientifico
Settore MED/23
Settore MEDS-13/C - Chirurgia cardiaca
English
Con Impact Factor ISI
Aortic valve replacement
Bioprostheses
Rapid-deployment aortic bioprostheses
Stented aortic bioprosthesis
D'Onofrio, A., Cibin, G., Tessari, C., Lorenzoni, G., Luzi, G., Manzan, E., et al. (2024). Multicenter, propensity-weighted comparison of stented, rapid-deployment and new-generation aortic valves. INTERNATIONAL JOURNAL OF CARDIOLOGY. HEART & VASCULATURE, 54 [10.1016/j.ijcha.2024.101487].
D'Onofrio, A; Cibin, G; Tessari, C; Lorenzoni, G; Luzi, G; Manzan, E; Gregori, D; Gerosa, G
Articolo su rivista
File in questo prodotto:
File Dimensione Formato  
main.pdf

accesso aperto

Tipologia: Versione Editoriale (PDF)
Licenza: Creative commons
Dimensione 728.5 kB
Formato Adobe PDF
728.5 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/397121
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 1
  • ???jsp.display-item.citation.isi??? 1
social impact