Objectives: Aortic stentless pericardial valves were introduced into clinical practice to combine properties of both stentless and pericardial prostheses. The aim of this single-center retrospective study was to assess midterm clinical and hemodynamic results of aortic valve replacement with the Sorin Pericarbon Freedom stentless bioprosthesis. Methods: From July 1999 through November 2005, 130 consecutive patients (73 [56.1%] male patients) underwent aortic valve replacement with the Sorin Pericarbon Freedom bioprosthesis at our institution. Mean age was 76 ± 5 years (range, 42-86 years), and associated procedures were performed in 50 (38.4%) patients; of these, 41 were coronary artery bypass grafts. Surgical intervention under urgent/emergency conditions and reoperations were performed in 18 (13.8%) and 7 (5.3%) patients, respectively. Mean crossclamp and cardiopulmonary bypass times were 82 ± 24 and 125 ± 40 minutes, respectively. All patients underwent clinical and echocardiographic follow-up (100% complete), and the total cumulative follow-up was 324 patient/years (mean, 2.5 ± 1.8; range, 6 months-7 years). Results: Overall hospital mortality was 8.4%. Overall patient survival was 63% ± 6% and 50% ± 10% at 5 and 7 years, respectively. Late deaths occurred in 23 patients, and 6 of them were valve related (1.8% patient/years). Freedom from valve-related death and reoperation was 91% ± 4% and 94% ± 4%, respectively, at 7 years. No structural valve deterioration was observed. Endocarditis, thromboembolism, and hemorrhagic complications occurred in 2 (0.6% patient/years), 1 (0.3% patient/years), and 1 (0.3% patient/years) patients, respectively. Mean transprosthetic gradients for valve sizes 23, 25, and 27 were 12.1 ± 3.8, 10.8 ± 3.8, and 9 ± 3.1 mm Hg, respectively. Conclusions: The Sorin Pericarbon Freedom stentless bioprosthesis provides good early and midterm results in terms of hemodynamic performance, survival, and freedom from valve-related complications. © 2007 The American Association for Thoracic Surgery.

D'Onofrio, A., Auriemma, S., Magagna, P., Favaro, A., Cannarella, A., Piccin, C., et al. (2007). Aortic valve replacement with the Sorin Pericarbon Freedom stentless prosthesis: 7 years{'} experience in 130 patients. JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 134(2), 491-495 [10.1016/j.jtcvs.2007.04.020].

Aortic valve replacement with the Sorin Pericarbon Freedom stentless prosthesis: 7 years{'} experience in 130 patients

D'Onofrio, A.
Writing – Original Draft Preparation
;
Favaro, A.;
2007-01-01

Abstract

Objectives: Aortic stentless pericardial valves were introduced into clinical practice to combine properties of both stentless and pericardial prostheses. The aim of this single-center retrospective study was to assess midterm clinical and hemodynamic results of aortic valve replacement with the Sorin Pericarbon Freedom stentless bioprosthesis. Methods: From July 1999 through November 2005, 130 consecutive patients (73 [56.1%] male patients) underwent aortic valve replacement with the Sorin Pericarbon Freedom bioprosthesis at our institution. Mean age was 76 ± 5 years (range, 42-86 years), and associated procedures were performed in 50 (38.4%) patients; of these, 41 were coronary artery bypass grafts. Surgical intervention under urgent/emergency conditions and reoperations were performed in 18 (13.8%) and 7 (5.3%) patients, respectively. Mean crossclamp and cardiopulmonary bypass times were 82 ± 24 and 125 ± 40 minutes, respectively. All patients underwent clinical and echocardiographic follow-up (100% complete), and the total cumulative follow-up was 324 patient/years (mean, 2.5 ± 1.8; range, 6 months-7 years). Results: Overall hospital mortality was 8.4%. Overall patient survival was 63% ± 6% and 50% ± 10% at 5 and 7 years, respectively. Late deaths occurred in 23 patients, and 6 of them were valve related (1.8% patient/years). Freedom from valve-related death and reoperation was 91% ± 4% and 94% ± 4%, respectively, at 7 years. No structural valve deterioration was observed. Endocarditis, thromboembolism, and hemorrhagic complications occurred in 2 (0.6% patient/years), 1 (0.3% patient/years), and 1 (0.3% patient/years) patients, respectively. Mean transprosthetic gradients for valve sizes 23, 25, and 27 were 12.1 ± 3.8, 10.8 ± 3.8, and 9 ± 3.1 mm Hg, respectively. Conclusions: The Sorin Pericarbon Freedom stentless bioprosthesis provides good early and midterm results in terms of hemodynamic performance, survival, and freedom from valve-related complications. © 2007 The American Association for Thoracic Surgery.
2007
Pubblicato
Rilevanza internazionale
Articolo
Comitato scientifico
Settore MED/23
Settore MEDS-13/C - Chirurgia cardiaca
English
D'Onofrio, A., Auriemma, S., Magagna, P., Favaro, A., Cannarella, A., Piccin, C., et al. (2007). Aortic valve replacement with the Sorin Pericarbon Freedom stentless prosthesis: 7 years{'} experience in 130 patients. JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 134(2), 491-495 [10.1016/j.jtcvs.2007.04.020].
D'Onofrio, A; Auriemma, S; Magagna, P; Favaro, A; Cannarella, A; Piccin, C; Bilotta, M; Abbiate, N; Lamascese, N; Fabbri, A
Articolo su rivista
File in questo prodotto:
File Dimensione Formato  
1-s2.0-S0022522307007660-main.pdf

solo utenti autorizzati

Tipologia: Versione Editoriale (PDF)
Licenza: Copyright dell'editore
Dimensione 190.92 kB
Formato Adobe PDF
190.92 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/397069
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 19
  • ???jsp.display-item.citation.isi??? 18
social impact