A multichannel microfluidic platform for real-time monitoring of epithelial barrier integrity by electrical impedance has been developed. Growth and polarization of human epithelial cells from the airway or gastrointestinal tract was continuously monitored over 5 days in 8 parallel, individually perfused microfluidic chips. Electrical impedance data were continuously recorded to monitor cell barrier formation using a low-cost bespoke impedance analyser. Data was analysed using an electric circuit model to extract the equivalent transepithelial electrical resistance and epithelial cell layer capacitance. The cell barrier integrity steadily increased overtime, achieving an average resistance of 418 ± 121 Ω cm2 (airway cells) or 207 ± 59 Ω cm2 (gastrointestinal cells) by day 5. The utility of the polarized airway epithelial barrier was demonstrated using a 24 hour challenge with double stranded RNA to mimic viral infection. This caused a rapid decrease in barrier integrity in association with disruption of tight junctions, whereas simultaneous treatment with a corticosteroid reduced this effect. The platform is able to measure barrier integrity in real-time and is scalable, thus has the potential to be used for drug development and testing.

Fernandes, J., Karra, N., Bowring, J., Reale, R., James, J., Blume, C., et al. (2022). Real-time monitoring of epithelial barrier function by impedance spectroscopy in a microfluidic platform. LAB ON A CHIP, 22(10), 2041-2054 [10.1039/d1lc01046h].

Real-time monitoring of epithelial barrier function by impedance spectroscopy in a microfluidic platform

Reale, Riccardo;
2022-01-01

Abstract

A multichannel microfluidic platform for real-time monitoring of epithelial barrier integrity by electrical impedance has been developed. Growth and polarization of human epithelial cells from the airway or gastrointestinal tract was continuously monitored over 5 days in 8 parallel, individually perfused microfluidic chips. Electrical impedance data were continuously recorded to monitor cell barrier formation using a low-cost bespoke impedance analyser. Data was analysed using an electric circuit model to extract the equivalent transepithelial electrical resistance and epithelial cell layer capacitance. The cell barrier integrity steadily increased overtime, achieving an average resistance of 418 ± 121 Ω cm2 (airway cells) or 207 ± 59 Ω cm2 (gastrointestinal cells) by day 5. The utility of the polarized airway epithelial barrier was demonstrated using a 24 hour challenge with double stranded RNA to mimic viral infection. This caused a rapid decrease in barrier integrity in association with disruption of tight junctions, whereas simultaneous treatment with a corticosteroid reduced this effect. The platform is able to measure barrier integrity in real-time and is scalable, thus has the potential to be used for drug development and testing.
2022
Pubblicato
Rilevanza internazionale
Articolo
Esperti anonimi
Settore ING-IND/34
Settore IBIO-01/A - Bioingegneria
English
Con Impact Factor ISI
Fernandes, J., Karra, N., Bowring, J., Reale, R., James, J., Blume, C., et al. (2022). Real-time monitoring of epithelial barrier function by impedance spectroscopy in a microfluidic platform. LAB ON A CHIP, 22(10), 2041-2054 [10.1039/d1lc01046h].
Fernandes, J; Karra, N; Bowring, J; Reale, R; James, J; Blume, C; Pell, Tj; Rowan, Wc; Davies, De; Swindle, Ej; Morgan, H
Articolo su rivista
File in questo prodotto:
File Dimensione Formato  
Fernandes_Morgan_LOC_2022.pdf

accesso aperto

Tipologia: Versione Editoriale (PDF)
Licenza: Creative commons
Dimensione 10.7 MB
Formato Adobe PDF
10.7 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/394080
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 11
  • ???jsp.display-item.citation.isi??? 10
social impact