background: we compare the accuracy of new intraoral scanners (IOSs) in full-arch digital implant impressions. methods: a master model with six scan bodies was milled in poly(methyl methacrylate), measured by using a coordinate measuring machine, and scanned 15 times with four IOSs: primescan, medit i500, Vatech EZ scan, and iTero. the software was developed to identify the position points on each scan body. The 3D position and distance analysis were performed. Results: The average and ± standard deviation of the 3D position analysis was 29 μm ± 6 μm for PrimeScan, 39 μm ± 6 μm for iTero, 48 μm ± 18 μm for mediti500, and 118 μm ± 24 μm for Vatech EZ scan (p < 0.05). conclusions: all IOSs are able to make a digital complete implant impression in vitro according to the average misfit value reported in literature (150 μm); however, the 3D distance analysis showed that only the primescan and iTero presented negligible systematic error sources.

Di Fiore, A., Graiff, L., Savio, G., Granata, S., Basilicata, M., Bollero, P., et al. (2022). Investigation of the Accuracy of Four Intraoral Scanners in Mandibular Full-Arch Digital Implant Impression: A Comparative In Vitro Study. INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH, 19(8), 1-6 [10.3390/ijerph19084719].

Investigation of the Accuracy of Four Intraoral Scanners in Mandibular Full-Arch Digital Implant Impression: A Comparative In Vitro Study

Bollero P.;
2022-01-01

Abstract

background: we compare the accuracy of new intraoral scanners (IOSs) in full-arch digital implant impressions. methods: a master model with six scan bodies was milled in poly(methyl methacrylate), measured by using a coordinate measuring machine, and scanned 15 times with four IOSs: primescan, medit i500, Vatech EZ scan, and iTero. the software was developed to identify the position points on each scan body. The 3D position and distance analysis were performed. Results: The average and ± standard deviation of the 3D position analysis was 29 μm ± 6 μm for PrimeScan, 39 μm ± 6 μm for iTero, 48 μm ± 18 μm for mediti500, and 118 μm ± 24 μm for Vatech EZ scan (p < 0.05). conclusions: all IOSs are able to make a digital complete implant impression in vitro according to the average misfit value reported in literature (150 μm); however, the 3D distance analysis showed that only the primescan and iTero presented negligible systematic error sources.
2022
Pubblicato
Rilevanza internazionale
Articolo
Esperti anonimi
Settore MEDS-16/A - Malattie odontostomatologiche
English
accuracy
CAD/CAM
dental implant
digital impression
full arch
intra-oral scanner
Di Fiore, A., Graiff, L., Savio, G., Granata, S., Basilicata, M., Bollero, P., et al. (2022). Investigation of the Accuracy of Four Intraoral Scanners in Mandibular Full-Arch Digital Implant Impression: A Comparative In Vitro Study. INTERNATIONAL JOURNAL OF ENVIRONMENTAL RESEARCH AND PUBLIC HEALTH, 19(8), 1-6 [10.3390/ijerph19084719].
Di Fiore, A; Graiff, L; Savio, G; Granata, S; Basilicata, M; Bollero, P; Meneghello, R
Articolo su rivista
File in questo prodotto:
File Dimensione Formato  
ijerph-19-04719.pdf

accesso aperto

Tipologia: Versione Editoriale (PDF)
Licenza: Creative commons
Dimensione 435.65 kB
Formato Adobe PDF
435.65 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/392029
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 11
  • ???jsp.display-item.citation.isi??? 11
social impact