introduction: the perceval sutureless biological prosthesis for aortic valve replacement has been introduced with the rationale for shortening surgical, extracorporeal circulation and aortic cross-clamping times, in order to reduce postoperative complications. aim: to evaluate early hemodynamic performance and immediate outcomes of implantation of the perceval sutureless bioprosthesis in comparison with the St. Jude trifecta sutured bioprosthesis for aortic valve replacement (Perfecta study). material and methods: between december 2014 and June 2023, 281 patients underwent St. Jude Trifecta implantation (n = 220, mean age: 75.2 ±6.5 years) and Perceval implantation, when indicated (n = 61, mean age: 77.9 ±5.1 years). concomitant CABG was performed in 73 (33%) and in 27 (44%) patients, respectively. results: extracorporeal circulation and cross-clamp times were significantly shorter in perceval patients in all aortic valve replacements (61 ±23 and 49 ±18 minutes vs. 96 ±36 and 67 ±21 minutes), and in isolated procedures (54 ±10 and 43 ±8 minutes vs. 84 ±28 and 66 ±21 minutes) (p < 0.0001, for all comparisons). operative mortality was absent and 2.7%, respectively (p = 0.2). postoperatively, low output cardiac syndrome (0% vs. 4.5%) and total rate of major cardiac and non-cardiac related complications (6.6% vs. 18.6%) were significantly lower in perceval patients (p = 0.01). echocardiography at discharge in comparison with preoperatively showed a relevant and similar decrease of mean and peak trans-aortic valve gradients for the trifecta prosthesis (11.6 ±4.3 vs. 50 ±15.2 mm Hg; 21.6 ±7.3 vs. 78.8 ±24 mm Hg) and for the perceval prosthesis (12.6 ±4.8 vs. 52 ±12.5 mm Hg; 22.6 ±7.9 vs. 77.8 ±16 mm Hg) (p < 0.00001, for all comparisons). Better global cardiac function was observed in perceval patients. concomitant multi-vessel and left main coronary artery disease (p = 0.046; HR = 4.6) and chronic pulmonary disease (p = 0.006; HR = 5.6) were detected as independent predictors of death and postoperative major complications. conclusions: early hemodynamic performance appears to be satisfactory with the use of trifecta sutured and perceval sutureless bioprostheses. perceval implantation allows reduction of surgical times, better preservation of myocardial contractile function and, consequently, reduction of the risk of postoperative complications.

Nardi, P., Altieri, C., Pisano, C., Buioni, D., Agneni, F., Grizzi, G., et al. (2024). Perceval sutureless bioprosthesis versus Trifecta sutured bioprosthesis for aortic valve replacement: immediate results of the Perfecta study. KARDIOCHIRURGIA I TORAKOCHIRURGIA POLSKA, 21(3), 153-161 [10.5114/kitp.2024.143499].

Perceval sutureless bioprosthesis versus Trifecta sutured bioprosthesis for aortic valve replacement: immediate results of the Perfecta study

Nardi, Paolo
Writing – Original Draft Preparation
;
Pisano, Calogera;Buioni, Dario;Agneni, Federico;Dakli, Martin;Scognamiglio, Mattia;Bassano, Carlo;Ruvolo, Giovanni
2024-01-01

Abstract

introduction: the perceval sutureless biological prosthesis for aortic valve replacement has been introduced with the rationale for shortening surgical, extracorporeal circulation and aortic cross-clamping times, in order to reduce postoperative complications. aim: to evaluate early hemodynamic performance and immediate outcomes of implantation of the perceval sutureless bioprosthesis in comparison with the St. Jude trifecta sutured bioprosthesis for aortic valve replacement (Perfecta study). material and methods: between december 2014 and June 2023, 281 patients underwent St. Jude Trifecta implantation (n = 220, mean age: 75.2 ±6.5 years) and Perceval implantation, when indicated (n = 61, mean age: 77.9 ±5.1 years). concomitant CABG was performed in 73 (33%) and in 27 (44%) patients, respectively. results: extracorporeal circulation and cross-clamp times were significantly shorter in perceval patients in all aortic valve replacements (61 ±23 and 49 ±18 minutes vs. 96 ±36 and 67 ±21 minutes), and in isolated procedures (54 ±10 and 43 ±8 minutes vs. 84 ±28 and 66 ±21 minutes) (p < 0.0001, for all comparisons). operative mortality was absent and 2.7%, respectively (p = 0.2). postoperatively, low output cardiac syndrome (0% vs. 4.5%) and total rate of major cardiac and non-cardiac related complications (6.6% vs. 18.6%) were significantly lower in perceval patients (p = 0.01). echocardiography at discharge in comparison with preoperatively showed a relevant and similar decrease of mean and peak trans-aortic valve gradients for the trifecta prosthesis (11.6 ±4.3 vs. 50 ±15.2 mm Hg; 21.6 ±7.3 vs. 78.8 ±24 mm Hg) and for the perceval prosthesis (12.6 ±4.8 vs. 52 ±12.5 mm Hg; 22.6 ±7.9 vs. 77.8 ±16 mm Hg) (p < 0.00001, for all comparisons). Better global cardiac function was observed in perceval patients. concomitant multi-vessel and left main coronary artery disease (p = 0.046; HR = 4.6) and chronic pulmonary disease (p = 0.006; HR = 5.6) were detected as independent predictors of death and postoperative major complications. conclusions: early hemodynamic performance appears to be satisfactory with the use of trifecta sutured and perceval sutureless bioprostheses. perceval implantation allows reduction of surgical times, better preservation of myocardial contractile function and, consequently, reduction of the risk of postoperative complications.
2024
Pubblicato
Rilevanza internazionale
Articolo
Comitato scientifico
Settore MED/23
Settore MEDS-13/C - Chirurgia cardiaca
English
Con Impact Factor ISI
surgical aortic valve replacement, sutureless prostheses, third-generation sutured biological prostheses.
Nardi, P., Altieri, C., Pisano, C., Buioni, D., Agneni, F., Grizzi, G., et al. (2024). Perceval sutureless bioprosthesis versus Trifecta sutured bioprosthesis for aortic valve replacement: immediate results of the Perfecta study. KARDIOCHIRURGIA I TORAKOCHIRURGIA POLSKA, 21(3), 153-161 [10.5114/kitp.2024.143499].
Nardi, P; Altieri, C; Pisano, C; Buioni, D; Agneni, F; Grizzi, G; Dakli, M; Salvati, Ac; Scognamiglio, M; Bassano, C; Ruvolo, G
Articolo su rivista
File in questo prodotto:
File Dimensione Formato  
Nardi Altieri Perfecta Study KiTP_Art_54884-10 2024.pdf

accesso aperto

Tipologia: Versione Editoriale (PDF)
Licenza: Creative commons
Dimensione 196.01 kB
Formato Adobe PDF
196.01 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/387072
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus ND
  • ???jsp.display-item.citation.isi??? ND
social impact