An improved thermal and fast neutrons detector is obtained, modifying a recently proposed multilayered homoepitaxial Chemical Vapor Deposition (CVD) diamond detector (M. Marinelli, et al., Appl. Phys. Lett. 89 (2006) 143509), where a <sup>6</sup>LiF layer deposited on the sensing layer was used to convert thermal neutrons into charged particles. By sandwiching this layer between two CVD diamond detectors connected in parallel, a better signal-to-background separation is achieved. This allows to use <sup>10</sup>B as converting element, so to realize a detector suitable for Boron Neutron Capture Therapy dosimetry. Also, the doubled detector volume enhances the sensitivity to fast neutrons. [All rights reserved Elsevier].

Almaviva, S., Marinelli, M., Milani, E., Prestopino, G., Tucciarone, A., Verona, C., et al. (2010). Improved performance in synthetic diamond neutron detectors: Application to boron neutron capture therapy. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT, 612(3), 580-582 [10.1016/j.nima.2009.08.016].

Improved performance in synthetic diamond neutron detectors: Application to boron neutron capture therapy

MARINELLI, MARCO;MILANI, ENRICO;TUCCIARONE, ALDO;Verona, C;VERONA RINATI, GIANLUCA;
2010-01-01

Abstract

An improved thermal and fast neutrons detector is obtained, modifying a recently proposed multilayered homoepitaxial Chemical Vapor Deposition (CVD) diamond detector (M. Marinelli, et al., Appl. Phys. Lett. 89 (2006) 143509), where a 6LiF layer deposited on the sensing layer was used to convert thermal neutrons into charged particles. By sandwiching this layer between two CVD diamond detectors connected in parallel, a better signal-to-background separation is achieved. This allows to use 10B as converting element, so to realize a detector suitable for Boron Neutron Capture Therapy dosimetry. Also, the doubled detector volume enhances the sensitivity to fast neutrons. [All rights reserved Elsevier].
gen-2010
Pubblicato
Rilevanza internazionale
Articolo
Sì, ma tipo non specificato
Settore FIS/01 - FISICA SPERIMENTALE
English
chemical vapour deposition;diamond;dosimeters;neutron capture therapy;semiconductor counters;
Almaviva, S., Marinelli, M., Milani, E., Prestopino, G., Tucciarone, A., Verona, C., et al. (2010). Improved performance in synthetic diamond neutron detectors: Application to boron neutron capture therapy. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT, 612(3), 580-582 [10.1016/j.nima.2009.08.016].
Almaviva, S; Marinelli, M; Milani, E; Prestopino, G; Tucciarone, A; Verona, C; VERONA RINATI, G; Angelone, M; Pillon, M
Articolo su rivista
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/35248
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 16
  • ???jsp.display-item.citation.isi??? 15
social impact