A synthetic single crystal diamond based dosimeter in a p-type/intrinsic/metal structure, operating in photovoltaic regime, is proposed for application in highly conformed radiotherapy dosimetry. The device was characterized by using 6 and 10 MV Bremsstrahlung X-ray beams and electron beams from 6 MeV up to 18 MeV, obtained by a CLINAC DHX Varian accelerator. All measurements were performed in a water phantom and commercial ionization chambers were used for calibration and comparison. Results showed a very good agreement of the diamond device response, as compared with the reference dosimeters, fast response times and high spatial resolution. One of such diamond dosimeters was then tested using a real Intensity Modulated Radiation Therapy (IMRT) prostate cancer treatment plan and its performance was compared with the ones from ionization chambers and a 2D diode array. The obtained results clearly assess the suitability of synthetic single crystal diamond for dose measurements in highly conformed radiotherapy and particularly in IMRT applications. [All rights reserved Elsevier].

Almaviva, S., Ciancaglioni, I., Consorti, R., De Notaristefani, F., Manfredotti, C., Marinelli, M., et al. (2010). Synthetic single crystal diamond dosimeters for conformal radiation therapy application. DIAMOND AND RELATED MATERIALS, 19(2-3), 217-220 [10.1016/j.diamond.2009.10.007].

Synthetic single crystal diamond dosimeters for conformal radiation therapy application

MARINELLI, MARCO;MILANI, ENRICO;Verona, C;VERONA RINATI, GIANLUCA
2010-02-01

Abstract

A synthetic single crystal diamond based dosimeter in a p-type/intrinsic/metal structure, operating in photovoltaic regime, is proposed for application in highly conformed radiotherapy dosimetry. The device was characterized by using 6 and 10 MV Bremsstrahlung X-ray beams and electron beams from 6 MeV up to 18 MeV, obtained by a CLINAC DHX Varian accelerator. All measurements were performed in a water phantom and commercial ionization chambers were used for calibration and comparison. Results showed a very good agreement of the diamond device response, as compared with the reference dosimeters, fast response times and high spatial resolution. One of such diamond dosimeters was then tested using a real Intensity Modulated Radiation Therapy (IMRT) prostate cancer treatment plan and its performance was compared with the ones from ionization chambers and a 2D diode array. The obtained results clearly assess the suitability of synthetic single crystal diamond for dose measurements in highly conformed radiotherapy and particularly in IMRT applications. [All rights reserved Elsevier].
feb-2010
Pubblicato
Rilevanza internazionale
Articolo
Sì, ma tipo non specificato
Settore FIS/01 - FISICA SPERIMENTALE
English
bremsstrahlung;diamond;dosimeters;ionisation chambers;radiation therapy;
Almaviva, S., Ciancaglioni, I., Consorti, R., De Notaristefani, F., Manfredotti, C., Marinelli, M., et al. (2010). Synthetic single crystal diamond dosimeters for conformal radiation therapy application. DIAMOND AND RELATED MATERIALS, 19(2-3), 217-220 [10.1016/j.diamond.2009.10.007].
Almaviva, S; Ciancaglioni, I; Consorti, R; De Notaristefani, F; Manfredotti, C; Marinelli, M; Milani, E; Petrucci, A; Prestopino, G; Verona, C; VERONA RINATI, G
Articolo su rivista
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/35247
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 12
  • ???jsp.display-item.citation.isi??? 10
social impact