A silicon detector and a diamond detector have been used at the SPS experimental North Area at CERN for diagnostics of beams of ultra-high energy lead ions (150 GeV/nucleon).The detectors are operated in pulse mode using a fast digital electronic chain. The discrimination capability of the two detectors is investigated using deposited energy and pulse shape analysis. It is shown that the detectors have good beam monitoring performances and can discriminate the main lead ion beam component from contamination of light and heavy fragments.
Cazzaniga, C., Kastriotou, M., García Alía, R., Fernandez-Martinez, P., Wyrwoll, V., Minniti, T., et al. (2021). Measurements of ultra-high energy lead ions using silicon and diamond detectors. NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH. SECTION A, ACCELERATORS, SPECTROMETERS, DETECTORS AND ASSOCIATED EQUIPMENT, 985 [10.1016/j.nima.2020.164671].
Measurements of ultra-high energy lead ions using silicon and diamond detectors
Minniti T.;
2021-01-01
Abstract
A silicon detector and a diamond detector have been used at the SPS experimental North Area at CERN for diagnostics of beams of ultra-high energy lead ions (150 GeV/nucleon).The detectors are operated in pulse mode using a fast digital electronic chain. The discrimination capability of the two detectors is investigated using deposited energy and pulse shape analysis. It is shown that the detectors have good beam monitoring performances and can discriminate the main lead ion beam component from contamination of light and heavy fragments.File | Dimensione | Formato | |
---|---|---|---|
Cazzaniga_Measurements-of-ultra-high-energy-lead-ions_2021.pdf
accesso aperto
Tipologia:
Versione Editoriale (PDF)
Licenza:
Creative commons
Dimensione
2.49 MB
Formato
Adobe PDF
|
2.49 MB | Adobe PDF | Visualizza/Apri |
I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.