This study deals with the thermal analysis of a transcutaneous energy transfer (TET) system, based on the magnetic resonant coupling technology, to wirelessly power a left ventricular assist device (LVAD). The transmitting planar coil is placed on the skin, just in front of the subcutaneously implanted receiving planar coil. The TET system permits to eliminate the infection due to the exit site of the percutaneous driveline cable that connects the LVAD with the external power supply and control system. Since the LVAD continuously requires high power (typically around 5 W as average value), the thermal aspects must be carefully kept under control. Aim of this paper is an extensive thermal analysis to properly design a suitable TET system.

Campi, T., Cruciani, S., Maradei, F., Montalto, A., Musumeci, F., Feliziani, M. (2022). Thermal Analysis of a Transcutaneous Energy Transfer System for a Left Ventricular Assist Device. IEEE JOURNAL OF ELECTROMAGNETICS, RF AND MICROWAVES IN MEDICINE AND BIOLOGY., 6(2), 253-259 [10.1109/JERM.2021.3109449].

Thermal Analysis of a Transcutaneous Energy Transfer System for a Left Ventricular Assist Device

Cruciani S.;
2022-01-01

Abstract

This study deals with the thermal analysis of a transcutaneous energy transfer (TET) system, based on the magnetic resonant coupling technology, to wirelessly power a left ventricular assist device (LVAD). The transmitting planar coil is placed on the skin, just in front of the subcutaneously implanted receiving planar coil. The TET system permits to eliminate the infection due to the exit site of the percutaneous driveline cable that connects the LVAD with the external power supply and control system. Since the LVAD continuously requires high power (typically around 5 W as average value), the thermal aspects must be carefully kept under control. Aim of this paper is an extensive thermal analysis to properly design a suitable TET system.
2022
Pubblicato
Rilevanza internazionale
Articolo
Esperti anonimi
Settore ING-IND/31 - ELETTROTECNICA
English
Heating systems
Biological tissues
Heart
Energy exchange
Electromagnetics
Couplings
Thermal analysis
Driveline infection (DLI)
left ventricular assist device (LVAD)
transcutaneous energy transfer (TET)
wireless power transfer (WPT)
Campi, T., Cruciani, S., Maradei, F., Montalto, A., Musumeci, F., Feliziani, M. (2022). Thermal Analysis of a Transcutaneous Energy Transfer System for a Left Ventricular Assist Device. IEEE JOURNAL OF ELECTROMAGNETICS, RF AND MICROWAVES IN MEDICINE AND BIOLOGY., 6(2), 253-259 [10.1109/JERM.2021.3109449].
Campi, T; Cruciani, S; Maradei, F; Montalto, A; Musumeci, F; Feliziani, M
Articolo su rivista
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/325203
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 2
  • ???jsp.display-item.citation.isi??? 1
social impact