This paper deals with the intrasystem electromagnetic interference (EMI) in a pacemaker equipped with a wireless power transfer (WPT) charging system. The WPT application to pacemakers is very new, and no results are yet published on possible EMI effects produced by the WPT coil currents in the pacemaker pacing leads. To this aim, an efficient and original co-simulation circuit/field method is proposed to predict the induced voltages on a pacing lead. In the numerical calculation, the pacemaker with WPT secondary coil and a pacing lead is implanted in a sophisticated human body model. The numerical results are validated by measurements showing a good agreement. The proposed approach would be a useful tool for pacemaker EMI compliance tests and safety assessment evaluations.

Campi, T., Cruciani, S., De Santis, V., Palandrani, F., Maradei, F., Feliziani, M. (2017). Induced Effects in a Pacemaker Equipped with a Wireless Power Transfer Charging System. IEEE TRANSACTIONS ON MAGNETICS, 53(6), 1-4 [10.1109/TMAG.2017.2661859].

Induced Effects in a Pacemaker Equipped with a Wireless Power Transfer Charging System

Cruciani S.;
2017-01-01

Abstract

This paper deals with the intrasystem electromagnetic interference (EMI) in a pacemaker equipped with a wireless power transfer (WPT) charging system. The WPT application to pacemakers is very new, and no results are yet published on possible EMI effects produced by the WPT coil currents in the pacemaker pacing leads. To this aim, an efficient and original co-simulation circuit/field method is proposed to predict the induced voltages on a pacing lead. In the numerical calculation, the pacemaker with WPT secondary coil and a pacing lead is implanted in a sophisticated human body model. The numerical results are validated by measurements showing a good agreement. The proposed approach would be a useful tool for pacemaker EMI compliance tests and safety assessment evaluations.
2017
Pubblicato
Rilevanza internazionale
Articolo
Esperti anonimi
Settore ING-IND/31 - ELETTROTECNICA
English
Con Impact Factor ISI
Coupling
electromagnetic compatibility
leads
pacemakers
transmission line (TL)
wireless power transfer
Campi, T., Cruciani, S., De Santis, V., Palandrani, F., Maradei, F., Feliziani, M. (2017). Induced Effects in a Pacemaker Equipped with a Wireless Power Transfer Charging System. IEEE TRANSACTIONS ON MAGNETICS, 53(6), 1-4 [10.1109/TMAG.2017.2661859].
Campi, T; Cruciani, S; De Santis, V; Palandrani, F; Maradei, F; Feliziani, M
Articolo su rivista
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/325110
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 18
  • ???jsp.display-item.citation.isi??? 17
social impact