This paper deals with the electric and magnetic field (EMF) safety aspects of a wireless power transfer (WPT) system based on magnetic resonant coupling between two coils. The primary coil is assumed to be on-body, while the secondary coil is assumed to be inside the human body and connected to a battery recharge system of an active implantable medical device such as a pacemaker. This study allows us to identify a good preliminary solution of the WPT coil configuration, compensation capacitor topology, and operational frequency. Demonstrative WPT systems operating at two different frequencies are proposed in order to verify the WPT performances. The EMF safety has been finally assessed by numerical dosimetry studies using anatomically realistic human body models revealing no particular concerns about this application.

Campi, T., Cruciani, S., Palandrani, F., De Santis, V., Hirata, A., Feliziani, M. (2016). Wireless power transfer charging system for AIMDs and pacemakers. IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 64(2), 633-642 [10.1109/TMTT.2015.2511011].

Wireless power transfer charging system for AIMDs and pacemakers

Cruciani S.;
2016-01-01

Abstract

This paper deals with the electric and magnetic field (EMF) safety aspects of a wireless power transfer (WPT) system based on magnetic resonant coupling between two coils. The primary coil is assumed to be on-body, while the secondary coil is assumed to be inside the human body and connected to a battery recharge system of an active implantable medical device such as a pacemaker. This study allows us to identify a good preliminary solution of the WPT coil configuration, compensation capacitor topology, and operational frequency. Demonstrative WPT systems operating at two different frequencies are proposed in order to verify the WPT performances. The EMF safety has been finally assessed by numerical dosimetry studies using anatomically realistic human body models revealing no particular concerns about this application.
2016
Pubblicato
Rilevanza internazionale
Articolo
Esperti anonimi
Settore ING-IND/31 - ELETTROTECNICA
English
Con Impact Factor ISI
Active implantable medical device (AIMD)
magnetic resonant coupling
electric and magnetic field (EMF) safety
numerical dosimetry
pacemaker
wireless power transfer (WPT)
Campi, T., Cruciani, S., Palandrani, F., De Santis, V., Hirata, A., Feliziani, M. (2016). Wireless power transfer charging system for AIMDs and pacemakers. IEEE TRANSACTIONS ON MICROWAVE THEORY AND TECHNIQUES, 64(2), 633-642 [10.1109/TMTT.2015.2511011].
Campi, T; Cruciani, S; Palandrani, F; De Santis, V; Hirata, A; Feliziani, M
Articolo su rivista
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/325106
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 211
  • ???jsp.display-item.citation.isi??? 170
social impact