An uninterruptable wireless power supply is developed for a left ventricular assist device (LVAD), which is an implanted, electrically driven centrifugal pump used to treat advanced heart failure pathologies. The LVADs are currently powered through a cable driveline (DL) that exits the patient's body through the skin. Here, an innovative uninterruptable power supply is presented, based on a hybrid wireless/wired architecture: the electric power of 10 W is transferred by the near field wireless power transfer technology operating at 300 kHz, from outside the human body to an implanted subcutaneous receiver, and then by an internal cable to the LVAD. A significant improvement of the subcutaneous receiver is here proposed, based on the integration in the receiving coil of a battery pack providing the following benefits: no-break power supply for the LVAD; peak load management; enhanced system reliability; elimination of DL infection; enhancement of patient's quality of life. The proposed system is first designed and optimized, then a demonstrator is made and tested in a laboratory using the latest LVAD generation. The measured coil currents are used as sources for dosimetric analysis to assess the compliance of the system with the electromagnetic field safety and thermal regulations.

Campi, T., Cruciani, S., Maradei, F., Montalto, A., Musumeci, F., Feliziani, M. (2021). Uninterruptable Transcutaneous Wireless Power Supply for an LVAD: Experimental Validation and EMF Safety Analysis. IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 63(5), 1717-1725 [10.1109/TEMC.2021.3102036].

Uninterruptable Transcutaneous Wireless Power Supply for an LVAD: Experimental Validation and EMF Safety Analysis

Cruciani S.;
2021-01-01

Abstract

An uninterruptable wireless power supply is developed for a left ventricular assist device (LVAD), which is an implanted, electrically driven centrifugal pump used to treat advanced heart failure pathologies. The LVADs are currently powered through a cable driveline (DL) that exits the patient's body through the skin. Here, an innovative uninterruptable power supply is presented, based on a hybrid wireless/wired architecture: the electric power of 10 W is transferred by the near field wireless power transfer technology operating at 300 kHz, from outside the human body to an implanted subcutaneous receiver, and then by an internal cable to the LVAD. A significant improvement of the subcutaneous receiver is here proposed, based on the integration in the receiving coil of a battery pack providing the following benefits: no-break power supply for the LVAD; peak load management; enhanced system reliability; elimination of DL infection; enhancement of patient's quality of life. The proposed system is first designed and optimized, then a demonstrator is made and tested in a laboratory using the latest LVAD generation. The measured coil currents are used as sources for dosimetric analysis to assess the compliance of the system with the electromagnetic field safety and thermal regulations.
2021
Pubblicato
Rilevanza internazionale
Articolo
Esperti anonimi
Settore ING-IND/31 - ELETTROTECNICA
English
Con Impact Factor ISI
Driveline infection (DLI)
electromagnetic field (EMF) safety
heart failure
left ventricular assist device (LVAD)
percutaneous driveline
transcutaneous energy transfer (TET)
wireless power transfer (WPT)
Campi, T., Cruciani, S., Maradei, F., Montalto, A., Musumeci, F., Feliziani, M. (2021). Uninterruptable Transcutaneous Wireless Power Supply for an LVAD: Experimental Validation and EMF Safety Analysis. IEEE TRANSACTIONS ON ELECTROMAGNETIC COMPATIBILITY, 63(5), 1717-1725 [10.1109/TEMC.2021.3102036].
Campi, T; Cruciani, S; Maradei, F; Montalto, A; Musumeci, F; Feliziani, M
Articolo su rivista
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/325105
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 2
  • ???jsp.display-item.citation.isi??? 2
social impact