OBJECTIVE:Radial artery occlusion is a potential complication of transradial procedures and its occurrence ranges from 0.8 to 30%. It is virtually always asymptomatic but the functional and sensorial consequences of a long acting hand hypoperfusion could go underestimated. CardioWaves is a novel photoplethysmograh device that allows us to detect the pulse wave amplitude of the blood flowing to the hand. Our objective was to assess in normal subjects the hand blood flow supplied by radial arteries and ulnopalmar arches, respectively, by using CardioWaves device during modified Allen's test (MAT). PATIENTS AND METHODS: MAT was performed on both hands of 60 normal subjects, age ranging 21 to 66 years, without any cardiovascular factor risk. RESULTS: Photoplethysmograh and MAT showed a high positive linear correlation (r=0.93). Despite that, MAT tends to give a higher reading by between 1.05 and 1.6 sec. 11 of 120 readings (9%) by CardioWaves showed values of radial/ulnar pulse amplitude ratio more than mean + 1 SD, suggesting a significant decrease in ulnopalmar arterial circulation when radial blood flow supply would ceased. CONCLUSIONS: The CardioWaves device allows us an accurate reading of the flow because of its independency from respiratory changes. Furthemore, the evaluation of radial and ulnar pulse wave amplitude and the ratio between them would reveal an insufficient blood flow supply by the ulnar artery irrespective of the MAT results. We suggest that their assessment before performing coronary angiography and interventions may reduce potential complication of transradial access.

Tanzilli, G., Truscelli, G., Barilla', F., Cocco, N., Pannitteri, G., Tanzilli, A., et al. (2015). Evaluation of hand circulation with cardiowaves photoplethysmograph device during allen test in healthy volunteers. EUROPEAN REVIEW FOR MEDICAL AND PHARMACOLOGICAL SCIENCES, 19(16), 3006-3011.

Evaluation of hand circulation with cardiowaves photoplethysmograph device during allen test in healthy volunteers

BARILLA', Francesco;
2015-01-01

Abstract

OBJECTIVE:Radial artery occlusion is a potential complication of transradial procedures and its occurrence ranges from 0.8 to 30%. It is virtually always asymptomatic but the functional and sensorial consequences of a long acting hand hypoperfusion could go underestimated. CardioWaves is a novel photoplethysmograh device that allows us to detect the pulse wave amplitude of the blood flowing to the hand. Our objective was to assess in normal subjects the hand blood flow supplied by radial arteries and ulnopalmar arches, respectively, by using CardioWaves device during modified Allen's test (MAT). PATIENTS AND METHODS: MAT was performed on both hands of 60 normal subjects, age ranging 21 to 66 years, without any cardiovascular factor risk. RESULTS: Photoplethysmograh and MAT showed a high positive linear correlation (r=0.93). Despite that, MAT tends to give a higher reading by between 1.05 and 1.6 sec. 11 of 120 readings (9%) by CardioWaves showed values of radial/ulnar pulse amplitude ratio more than mean + 1 SD, suggesting a significant decrease in ulnopalmar arterial circulation when radial blood flow supply would ceased. CONCLUSIONS: The CardioWaves device allows us an accurate reading of the flow because of its independency from respiratory changes. Furthemore, the evaluation of radial and ulnar pulse wave amplitude and the ratio between them would reveal an insufficient blood flow supply by the ulnar artery irrespective of the MAT results. We suggest that their assessment before performing coronary angiography and interventions may reduce potential complication of transradial access.
2015
Pubblicato
Rilevanza internazionale
Articolo
Esperti anonimi
Settore MED/11 - MALATTIE DELL'APPARATO CARDIOVASCOLARE
English
modified allen's test
photoplethysmograph
hand circulation
radial arteries
ulnopalmar arterial arch
Tanzilli, G., Truscelli, G., Barilla', F., Cocco, N., Pannitteri, G., Tanzilli, A., et al. (2015). Evaluation of hand circulation with cardiowaves photoplethysmograph device during allen test in healthy volunteers. EUROPEAN REVIEW FOR MEDICAL AND PHARMACOLOGICAL SCIENCES, 19(16), 3006-3011.
Tanzilli, G; Truscelli, G; Barilla', F; Cocco, N; Pannitteri, G; Tanzilli, A; Al Kindy, S; Mangieri, E; Gaudio, C
Articolo su rivista
File in questo prodotto:
File Dimensione Formato  
Tanzilli_Evaluation_2015.pdf

solo utenti autorizzati

Dimensione 278.52 kB
Formato Adobe PDF
278.52 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/290827
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 6
  • ???jsp.display-item.citation.isi??? 5
social impact