Abstract Background: The purpose of the in vitro study is to investigate and compare the morphological features and the chemical stability in weight of two different polyurethane-based blends, Smart Track (LD30) and Exceed30 (EX30), used for orthodontic aligners manufacture before and after the oral usage. Methods: Twenty orthodontic aligners were randomly selected: 10 LD30 and 10 EX30, each group was divided in two subgroups, never used and intra-orally aged. By the employment of a Stereomicroscope, a section of 5 × 5 mm was cut from the buccal surface of the incisal region of each aligner. All samples were subjected to Scanning Electron Microscopy and Ageing tests in different solutions to simulate the hostility of the oral environment. The statistical method used was t-test. Results: At SEM images, LD30 appears more homogeneous in texture respect to EX30. However, after clinical usage, both materials show significant structural alterations: findings have been supported by higher magnifications at SEM, by which it is clearly to observe many superficial cracks cross through the polymer structures of LD30U, absent in never used samples. LD30U surface becomes also smoother due to the disappearance of most of the conglomerates, but at the same time also rougher while EX30U shows a greater irregularity and porosity in which large and deep cracks are also highlighted. Although these changes occur persistently, in the aging tests no significant weight loss from both materials has been found, confirming the initial hypothesis of a good chemical stability and safety of both polyurethane mixtures even in conditions of severe hostility. Conclusion: LD30 is the expression of the technological evolution of EX30, this is made evident above all by its morphological architecture, more homogeneous and defined but also by the chemical stability that can be appreciated even in evident critic situations.

Condo', R., Mampieri, G., Giancotti, A., Cerroni, L., Pasquantonio, G., Divizia, A., et al. (2021). SEM characterization and ageing analysis on two generation of invisible aligners. BMC ORAL HEALTH, 21(1) [10.1186/s12903-021-01676-z].

SEM characterization and ageing analysis on two generation of invisible aligners

Roberta Condò
Writing – Review & Editing
;
Gianluca Mampieri;Aldo Giancotti;Loredana Cerroni;Guido Pasquantonio;Barbara Mecheri;
2021-06-01

Abstract

Abstract Background: The purpose of the in vitro study is to investigate and compare the morphological features and the chemical stability in weight of two different polyurethane-based blends, Smart Track (LD30) and Exceed30 (EX30), used for orthodontic aligners manufacture before and after the oral usage. Methods: Twenty orthodontic aligners were randomly selected: 10 LD30 and 10 EX30, each group was divided in two subgroups, never used and intra-orally aged. By the employment of a Stereomicroscope, a section of 5 × 5 mm was cut from the buccal surface of the incisal region of each aligner. All samples were subjected to Scanning Electron Microscopy and Ageing tests in different solutions to simulate the hostility of the oral environment. The statistical method used was t-test. Results: At SEM images, LD30 appears more homogeneous in texture respect to EX30. However, after clinical usage, both materials show significant structural alterations: findings have been supported by higher magnifications at SEM, by which it is clearly to observe many superficial cracks cross through the polymer structures of LD30U, absent in never used samples. LD30U surface becomes also smoother due to the disappearance of most of the conglomerates, but at the same time also rougher while EX30U shows a greater irregularity and porosity in which large and deep cracks are also highlighted. Although these changes occur persistently, in the aging tests no significant weight loss from both materials has been found, confirming the initial hypothesis of a good chemical stability and safety of both polyurethane mixtures even in conditions of severe hostility. Conclusion: LD30 is the expression of the technological evolution of EX30, this is made evident above all by its morphological architecture, more homogeneous and defined but also by the chemical stability that can be appreciated even in evident critic situations.
giu-2021
Pubblicato
Rilevanza internazionale
Articolo
Esperti anonimi
Settore MED/28 - MALATTIE ODONTOSTOMATOLOGICHE
English
Con Impact Factor ISI
Ageing tests
Invisalign® aligner
Polymer
Scanning electron microscopy
Condo', R., Mampieri, G., Giancotti, A., Cerroni, L., Pasquantonio, G., Divizia, A., et al. (2021). SEM characterization and ageing analysis on two generation of invisible aligners. BMC ORAL HEALTH, 21(1) [10.1186/s12903-021-01676-z].
Condo', R; Mampieri, G; Giancotti, A; Cerroni, L; Pasquantonio, G; Divizia, A; Convertino, A; Mecheri, B; Maiolo, L
Articolo su rivista
File in questo prodotto:
File Dimensione Formato  
s12903-021-01676-z.pdf

accesso aperto

Descrizione: Articolo principale
Tipologia: Versione Editoriale (PDF)
Licenza: Copyright dell'editore
Dimensione 2.27 MB
Formato Adobe PDF
2.27 MB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/283089
Citazioni
  • ???jsp.display-item.citation.pmc??? 5
  • Scopus 10
  • ???jsp.display-item.citation.isi??? 16
social impact