Objectives: The aim of this retrospective multicenter study was to compare early clinical and hemodynamic outcomes of Perceval-S sutureless (Livanova, London, United Kingdom) and Intuity rapid-deployment (Edwards Lifesciences, Irvine, Calif) bioprostheses.Methods: Data from patients who underwent isolated or combined aortic valve replacement with Perceval-S and with Intuity bioprostheses at 18 cardiac surgical institutions were analyzed. Propensity matching was performed to identify similar patient cohorts.Results: We included 911 patients from March 2011 until May 2017. Perceval-S and Intuity valves were implanted in 349 (38.3%) and in 562 (61.7%) patients, respectively. Propensity score identified 117 matched pairs. In the matched cohort, device success was 99.1% and 100% in Perceval-S and Intuity group, respectively (P = 1.000). Thirty-day Valve Academic Research Consortium mortality occurred in 2 (1.7%) and 4 (3.4%) patients in the Perceval-S and in Intuity group, respectively (P = .6834). The rate of postoperative new permanent pacemaker implantation was 6%(7 patients) and 6.8%(8 patients) in the Perceval-S and in Intuity group, respectively (P = .7896). Perceval-S valve implantation requires significantly shorter aortic crossclamp and cardiopulmonary bypass times than Intuity valve implantation (aortic crossclamp time for isolated, 52 +/- 14 minutes vs 62 +/- 24 minutes; P<.0001). Peak transaortic gradients were 22.4 +/- 8.1 mm Hg and 19.6 +/- 6.7 mm Hg (P = .0144), whereas mean gradients were 11.8 +/- 4.7 mm Hg and 10.5 +/- 3.9 mm Hg (P = .0388) in the Perceval-S and Intuity groups, respectively.Conclusions: Sutureless Perceval-S and rapid-deployment Intuity bioprostheses provide good and similar early clinical and hemodynamic outcomes. Perceval-S valve implantation requires shorter crossclamp and cardiopulmonary bypass times, whereas Intuity valve implantation provides lower transaortic peak and mean gradients.

D'Onofrio, A., Salizzoni, S., Filippini, C., Tessari, C., Bagozzi, L., Messina, A., et al. (2019). Surgical aortic valve replacement with new-generation bioprostheses: Sutureless versus rapid-deployment. THE JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 159(2), 432-442.e1 [10.1016/j.jtcvs.2019.02.135].

Surgical aortic valve replacement with new-generation bioprostheses: Sutureless versus rapid-deployment

De Paulis R.;
2019-05-01

Abstract

Objectives: The aim of this retrospective multicenter study was to compare early clinical and hemodynamic outcomes of Perceval-S sutureless (Livanova, London, United Kingdom) and Intuity rapid-deployment (Edwards Lifesciences, Irvine, Calif) bioprostheses.Methods: Data from patients who underwent isolated or combined aortic valve replacement with Perceval-S and with Intuity bioprostheses at 18 cardiac surgical institutions were analyzed. Propensity matching was performed to identify similar patient cohorts.Results: We included 911 patients from March 2011 until May 2017. Perceval-S and Intuity valves were implanted in 349 (38.3%) and in 562 (61.7%) patients, respectively. Propensity score identified 117 matched pairs. In the matched cohort, device success was 99.1% and 100% in Perceval-S and Intuity group, respectively (P = 1.000). Thirty-day Valve Academic Research Consortium mortality occurred in 2 (1.7%) and 4 (3.4%) patients in the Perceval-S and in Intuity group, respectively (P = .6834). The rate of postoperative new permanent pacemaker implantation was 6%(7 patients) and 6.8%(8 patients) in the Perceval-S and in Intuity group, respectively (P = .7896). Perceval-S valve implantation requires significantly shorter aortic crossclamp and cardiopulmonary bypass times than Intuity valve implantation (aortic crossclamp time for isolated, 52 +/- 14 minutes vs 62 +/- 24 minutes; P<.0001). Peak transaortic gradients were 22.4 +/- 8.1 mm Hg and 19.6 +/- 6.7 mm Hg (P = .0144), whereas mean gradients were 11.8 +/- 4.7 mm Hg and 10.5 +/- 3.9 mm Hg (P = .0388) in the Perceval-S and Intuity groups, respectively.Conclusions: Sutureless Perceval-S and rapid-deployment Intuity bioprostheses provide good and similar early clinical and hemodynamic outcomes. Perceval-S valve implantation requires shorter crossclamp and cardiopulmonary bypass times, whereas Intuity valve implantation provides lower transaortic peak and mean gradients.
mag-2019
Online ahead of print
Rilevanza internazionale
Articolo
Sì, ma tipo non specificato
Settore MED/23 - CHIRURGIA CARDIACA
English
aortic valve; rapid deployment aortic bioprosthesis; replacement; sutureless aortic bioprosthesis; Aged; Aged, 80 and over; Aortic Valve; Aortic Valve Stenosis; Female; Heart Valve Prosthesis Implantation; Male; Retrospective Studies; Sutureless Surgical Procedures; Bioprosthesis; Heart Valve Prosthesis
D'Onofrio, A., Salizzoni, S., Filippini, C., Tessari, C., Bagozzi, L., Messina, A., et al. (2019). Surgical aortic valve replacement with new-generation bioprostheses: Sutureless versus rapid-deployment. THE JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 159(2), 432-442.e1 [10.1016/j.jtcvs.2019.02.135].
D'Onofrio, A; Salizzoni, S; Filippini, C; Tessari, C; Bagozzi, L; Messina, A; Troise, G; Tomba, Md; Rambaldini, M; Dalen, M; Alamanni, F; Massetti, M; Mignosa, C; Russo, C; Salvador, L; Di Bartolomeo, R; Maselli, D; De Paulis, R; Alfieri, O; De Filippo, Cm; Portoghese, M; Bortolotti, U; Rinaldi, M; Gerosa, G
Articolo su rivista
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/268241
Citazioni
  • ???jsp.display-item.citation.pmc??? 14
  • Scopus 29
  • ???jsp.display-item.citation.isi??? 24
social impact