The International Federation of Clinical Chemistry and Laboratory Medicine has introduced in recent times the turnaround time (TAT) as mandatory quality indicator for the postanalytical phase. Classic TAT indicators, namely, average, median, 90th percentile and proportion of acceptable test (PAT), are in use since almost 40 years and to date represent the mainstay for gauging the laboratory timeliness. In this study, we investigated the performance of the Six Sigma Z-score, which was previously introduced as a device for the quantitative assessment of timeliness.
Ialongo, C., Bernardini, S. (2018). Validation of the Six Sigma Z-score for the quality assessment of clinical laboratory timeliness. CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 56(4), 595-601 [10.1515/cclm-2017-0642].
Validation of the Six Sigma Z-score for the quality assessment of clinical laboratory timeliness
Bernardini S.
Supervision
2018-01-01
Abstract
The International Federation of Clinical Chemistry and Laboratory Medicine has introduced in recent times the turnaround time (TAT) as mandatory quality indicator for the postanalytical phase. Classic TAT indicators, namely, average, median, 90th percentile and proportion of acceptable test (PAT), are in use since almost 40 years and to date represent the mainstay for gauging the laboratory timeliness. In this study, we investigated the performance of the Six Sigma Z-score, which was previously introduced as a device for the quantitative assessment of timeliness.I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.