Technologies enabling in vivo range verification during proton therapy are actively sought as a means to reduce the clinical safety margins currently adopted to avoid tumor underdosage. In this contribution, we applied the semi-empirical theory of radiation-induced vaporization of superheated liquids to coated nanodroplets. Nanodroplets are injectable phase-change contrast agents that can vaporize into highly echogenic microbubbles to provide contrast in ultrasound images. We exposed nanodroplet dispersions in aqueous phantoms to monoenergetic proton beams of varying energies and doses. Ultrasound imaging of the phantoms revealed that radiation-induced droplet vaporization occurred in regions proximal to the proton Bragg peak. A statistically significant increase in contrast was observed in irradiated regions for doses as low as 2 Gy and found to be proportional to the proton fluence. The absence of enhanced response in the vicinity of the Bragg peak, combined with theoretical considerations, suggest that droplet vaporization is induced by high linear energy transfer (LET) recoil ions produced by nuclear reactions with incoming protons. Vaporization profiles were compared to non-elastic cross sections and LET characteristics of oxygen recoils. Shifts between the ultrasound image contrast drop and the expected proton range showed a sub-millimeter reproducibility. These early findings confirm the potential of superheated nanodroplets as a novel tool for proton range verification.

Carlier, B., Heymans, S., Noojiens, S., Toumia, Y., Igram, M., Paradossi, G., et al. (2020). Proton range verification with ultrasound imaging using injectable radiation sensitive nanodroplets: a feasibility study. PHYSICS IN MEDICINE AND BIOLOGY, 65(6) [10.1088/1361-6560/ab7506].

Proton range verification with ultrasound imaging using injectable radiation sensitive nanodroplets: a feasibility study

Toumia, Y;Paradossi, G;
2020-02-14

Abstract

Technologies enabling in vivo range verification during proton therapy are actively sought as a means to reduce the clinical safety margins currently adopted to avoid tumor underdosage. In this contribution, we applied the semi-empirical theory of radiation-induced vaporization of superheated liquids to coated nanodroplets. Nanodroplets are injectable phase-change contrast agents that can vaporize into highly echogenic microbubbles to provide contrast in ultrasound images. We exposed nanodroplet dispersions in aqueous phantoms to monoenergetic proton beams of varying energies and doses. Ultrasound imaging of the phantoms revealed that radiation-induced droplet vaporization occurred in regions proximal to the proton Bragg peak. A statistically significant increase in contrast was observed in irradiated regions for doses as low as 2 Gy and found to be proportional to the proton fluence. The absence of enhanced response in the vicinity of the Bragg peak, combined with theoretical considerations, suggest that droplet vaporization is induced by high linear energy transfer (LET) recoil ions produced by nuclear reactions with incoming protons. Vaporization profiles were compared to non-elastic cross sections and LET characteristics of oxygen recoils. Shifts between the ultrasound image contrast drop and the expected proton range showed a sub-millimeter reproducibility. These early findings confirm the potential of superheated nanodroplets as a novel tool for proton range verification.
14-feb-2020
Pubblicato
Rilevanza internazionale
Articolo
Esperti anonimi
Settore CHIM/02 - CHIMICA FISICA
English
Con Impact Factor ISI
ultrasound
nanodroplets
proton therapy
range verification
dosimetry
Carlier, B., Heymans, S., Noojiens, S., Toumia, Y., Igram, M., Paradossi, G., et al. (2020). Proton range verification with ultrasound imaging using injectable radiation sensitive nanodroplets: a feasibility study. PHYSICS IN MEDICINE AND BIOLOGY, 65(6) [10.1088/1361-6560/ab7506].
Carlier, B; Heymans, S; Noojiens, S; Toumia, Y; Igram, M; Paradossi, G; D'Agostino, E; Himmelreich, U; D'Hooge, J; Van Den Abeele, K; Sterpin, E
Articolo su rivista
File in questo prodotto:
File Dimensione Formato  
Carlier_2020_Phys._Med._Biol._65_065013.pdf

solo utenti autorizzati

Tipologia: Versione Editoriale (PDF)
Licenza: Copyright dell'editore
Dimensione 69.8 MB
Formato Adobe PDF
69.8 MB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/230163
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 24
  • ???jsp.display-item.citation.isi??? 24
social impact