Purpose Sleep-related breathing disorders (SRBD) have been associated with adverse cardiovascular events and prognosis. Some modern pacemakers (PMK) and implantable cardioverter defibrillators (ICD) are equipped with thoracic impedance monitoring systems which allow detecting fluctuations in tidal volume occurring in SRBD. This study aimed at assessing the accuracy of a novel transthoracic impedance-based PMK/ICD sensor for automatic detection of SRBD. Methods We prospectively enrolled 31 consecutive patients scheduled for implantation of a PMK or an ICD with sleep respiratory disturbances monitoring capability. The results of unattended home nocturnal recordings expressed as apnea/hypopnea index (AHI) were compared with the respiratory disturbance index (RDI) stored in the cardiac device memory at the time polygraphy was performed. Results Among the 31 enrolled patients (20 men, mean age 70 +/- 8 years), a comparison between the data from polygraphy and cardiac device was feasible in 29 patients. According to the results of home nocturnal recording, a severe SRBD was observed in 7/29 (24.1%) patients. On the basis of receiver-operating characteristic curve analysis of RDI values, the optimal RDI cutoff value to identify severe SRBD was 47 episodes/h (sensitivity 100%, specificity 100%). Bland-Altman agreement analysis of AHI-RDI revealed a bias between measurements of - 11 episodes/h, with limits of agreement - 38 to 15 episodes/h. Conclusions In the present study, the novel transthoracic impedance-based monitoring system ApneaScan (TM) appeared effective in screening PMK and ICD patients for SRBD.

Della Rocca, D.g., Albanese, M., Placidi, F., Forle, G.b., Di Biase, L., Ribatti, V., et al. (2019). Feasibility of automated detection of sleep apnea using implantable pacemakers and defibrillators: a comparison with simultaneous polysomnography recording. JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 56(3), 327-333 [10.1007/s10840-019-00631-x].

Feasibility of automated detection of sleep apnea using implantable pacemakers and defibrillators: a comparison with simultaneous polysomnography recording

Albanese M.;Placidi F.;Ribatti V.;Santini L.;Izzi F.;Natale A.;Mercuri N. B.;
2019-01-01

Abstract

Purpose Sleep-related breathing disorders (SRBD) have been associated with adverse cardiovascular events and prognosis. Some modern pacemakers (PMK) and implantable cardioverter defibrillators (ICD) are equipped with thoracic impedance monitoring systems which allow detecting fluctuations in tidal volume occurring in SRBD. This study aimed at assessing the accuracy of a novel transthoracic impedance-based PMK/ICD sensor for automatic detection of SRBD. Methods We prospectively enrolled 31 consecutive patients scheduled for implantation of a PMK or an ICD with sleep respiratory disturbances monitoring capability. The results of unattended home nocturnal recordings expressed as apnea/hypopnea index (AHI) were compared with the respiratory disturbance index (RDI) stored in the cardiac device memory at the time polygraphy was performed. Results Among the 31 enrolled patients (20 men, mean age 70 +/- 8 years), a comparison between the data from polygraphy and cardiac device was feasible in 29 patients. According to the results of home nocturnal recording, a severe SRBD was observed in 7/29 (24.1%) patients. On the basis of receiver-operating characteristic curve analysis of RDI values, the optimal RDI cutoff value to identify severe SRBD was 47 episodes/h (sensitivity 100%, specificity 100%). Bland-Altman agreement analysis of AHI-RDI revealed a bias between measurements of - 11 episodes/h, with limits of agreement - 38 to 15 episodes/h. Conclusions In the present study, the novel transthoracic impedance-based monitoring system ApneaScan (TM) appeared effective in screening PMK and ICD patients for SRBD.
2019
Pubblicato
Rilevanza internazionale
Articolo
Esperti anonimi
Settore MED/26 - NEUROLOGIA
English
Implantable defibrillator; Pacemaker; Sleep apnea; Sleep apnea screening; Transthoracic impedance
Della Rocca, D.g., Albanese, M., Placidi, F., Forle, G.b., Di Biase, L., Ribatti, V., et al. (2019). Feasibility of automated detection of sleep apnea using implantable pacemakers and defibrillators: a comparison with simultaneous polysomnography recording. JOURNAL OF INTERVENTIONAL CARDIAC ELECTROPHYSIOLOGY, 56(3), 327-333 [10.1007/s10840-019-00631-x].
Della Rocca, Dg; Albanese, M; Placidi, F; Forle, Gb; Di Biase, L; Ribatti, V; Santini, L; Izzi, F; Cicchini, L; Lovecchio, M; Valsecchi, S; Lavalle, C; Natale, A; Mercuri, Nb; Romigi, A
Articolo su rivista
File in questo prodotto:
File Dimensione Formato  
DellaRocca2019_Article_FeasibilityOfAutomatedDetectio(1).pdf

accesso aperto

Licenza: Copyright dell'editore
Dimensione 778.22 kB
Formato Adobe PDF
778.22 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/228741
Citazioni
  • ???jsp.display-item.citation.pmc??? 7
  • Scopus 11
  • ???jsp.display-item.citation.isi??? 9
social impact