Following international guidelines by the IEEE and a well-established calibration procedure developed at the Hungarian Institute of Isotopes in Budapest, the experimental absolute efficiency calibration of an Ortec n-type coaxial high-purity germanium (HPGe) detector (GMX Profile series) has been performed, in view of direct application to quantitative PGAA and T-PGAA investigations of Cultural Heritage multicomponent manufacts.
Parmentier, A., Arcidiacono, L., Senesi, R., Romanelli, G., Andreani, C., Moir, J., et al. (2018). Absolute efficiency calibration of a coaxial HPGe detector for quantitative PGAA and T-PGAA. JOURNAL OF PHYSICS. CONFERENCE SERIES, 1055(1), 012010 [10.1088/1742-6596/1055/1/012010].
Absolute efficiency calibration of a coaxial HPGe detector for quantitative PGAA and T-PGAA
Parmentier A.;Senesi R.;Romanelli G.;Andreani C.;Festa G.
2018-01-01
Abstract
Following international guidelines by the IEEE and a well-established calibration procedure developed at the Hungarian Institute of Isotopes in Budapest, the experimental absolute efficiency calibration of an Ortec n-type coaxial high-purity germanium (HPGe) detector (GMX Profile series) has been performed, in view of direct application to quantitative PGAA and T-PGAA investigations of Cultural Heritage multicomponent manufacts.File | Dimensione | Formato | |
---|---|---|---|
184.pdf
accesso aperto
Licenza:
Non specificato
Dimensione
617.4 kB
Formato
Adobe PDF
|
617.4 kB | Adobe PDF | Visualizza/Apri |