Purpose In this work, the LET-dependence of the response of synthetic diamond detectors is investigated in different particle beams. Method Measurements were performed in three nonmodulated particle beams (proton, carbon, and oxygen). The response of five synthetic diamond detectors was compared to the response of a Markus or an Advanced Markus ionization chamber. The synthetic diamond detectors were used with their axis parallel to the beam axis and without any bias voltage. A high bias voltage was applied to the ionization chambers, to minimize ion recombination, for which no correction is applied (+300 V and +400 V were applied to the Markus and Advanced Markus ionization chambers respectively). Results The ratio between the normalized response of the synthetic diamond detectors and the normalized response of the ionization chamber shows an under-response of the synthetic diamond detectors in carbon and oxygen ion beams. No under-response of the synthetic diamond detectors is observed in protons. For each beam, combining results obtained for the five synthetic diamond detectors and considering the uncertainties, a linear fit of the ratio between the normalized response of the synthetic diamond detectors and the normalized response of the ionization chamber is determined. The response of the synthetic diamond detectors can be described as a function of LET as (−6.22E-4 ± 3.17E-3) • LET + (0.99 ± 0.01) in proton beam, (−2.51E-4 ± 1.18E-4) • LET + (1.01 ± 0.01) in carbon ion beam and (−2.77E-4 ± 0.56E-4) • LET + (1.03 ± 0.01) in oxygen ion beam. Combining results obtained in carbon and oxygen ion beams, a LET dependence of about 0.026% (±0.013%) per keV/μm is estimated. Conclusions Due to the high LET value, a LET dependence of the response of the synthetic diamond detector was observed in the case of carbon and oxygen beams. The effect was found to be negligible in proton beams, due to the low LET value. The under-response of the synthetic diamond detector may result from the recombination of electron/hole in the thin synthetic diamond layer, due to the high LET-values. More investigations are required to confirm this assumption.In this work, the LET-dependence of the response of synthetic diamond detectors is investigated in different particle beams.

Rossomme, S., Marinelli, M., Verona-Rinati, G., Romano, F., Cirrone, P., Kacperek, A., et al. (2017). Response of synthetic diamond detectors in proton, carbon, and oxygen ion beams. MEDICAL PHYSICS, 44(10), 5445-5449 [10.1002/mp.12473].

Response of synthetic diamond detectors in proton, carbon, and oxygen ion beams

Marinelli M.;Verona-Rinati G.;
2017-01-01

Abstract

Purpose In this work, the LET-dependence of the response of synthetic diamond detectors is investigated in different particle beams. Method Measurements were performed in three nonmodulated particle beams (proton, carbon, and oxygen). The response of five synthetic diamond detectors was compared to the response of a Markus or an Advanced Markus ionization chamber. The synthetic diamond detectors were used with their axis parallel to the beam axis and without any bias voltage. A high bias voltage was applied to the ionization chambers, to minimize ion recombination, for which no correction is applied (+300 V and +400 V were applied to the Markus and Advanced Markus ionization chambers respectively). Results The ratio between the normalized response of the synthetic diamond detectors and the normalized response of the ionization chamber shows an under-response of the synthetic diamond detectors in carbon and oxygen ion beams. No under-response of the synthetic diamond detectors is observed in protons. For each beam, combining results obtained for the five synthetic diamond detectors and considering the uncertainties, a linear fit of the ratio between the normalized response of the synthetic diamond detectors and the normalized response of the ionization chamber is determined. The response of the synthetic diamond detectors can be described as a function of LET as (−6.22E-4 ± 3.17E-3) • LET + (0.99 ± 0.01) in proton beam, (−2.51E-4 ± 1.18E-4) • LET + (1.01 ± 0.01) in carbon ion beam and (−2.77E-4 ± 0.56E-4) • LET + (1.03 ± 0.01) in oxygen ion beam. Combining results obtained in carbon and oxygen ion beams, a LET dependence of about 0.026% (±0.013%) per keV/μm is estimated. Conclusions Due to the high LET value, a LET dependence of the response of the synthetic diamond detector was observed in the case of carbon and oxygen beams. The effect was found to be negligible in proton beams, due to the low LET value. The under-response of the synthetic diamond detector may result from the recombination of electron/hole in the thin synthetic diamond layer, due to the high LET-values. More investigations are required to confirm this assumption.In this work, the LET-dependence of the response of synthetic diamond detectors is investigated in different particle beams.
2017
Pubblicato
Rilevanza internazionale
Articolo
Esperti anonimi
Settore FIS/07 - FISICA APPLICATA (A BENI CULTURALI, AMBIENTALI, BIOLOGIA E MEDICINA)
English
LET-dependence; particle beams; relative dosimetry; synthetic diamond detector
Rossomme, S., Marinelli, M., Verona-Rinati, G., Romano, F., Cirrone, P., Kacperek, A., et al. (2017). Response of synthetic diamond detectors in proton, carbon, and oxygen ion beams. MEDICAL PHYSICS, 44(10), 5445-5449 [10.1002/mp.12473].
Rossomme, S; Marinelli, M; Verona-Rinati, G; Romano, F; Cirrone, Pag; Kacperek, A; Vynckier, S; Palmans, H
Articolo su rivista
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/193413
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 11
  • ???jsp.display-item.citation.isi??? 10
social impact