BACKGROUND: The use of dried blood spots (DBS) for HIV-1 viral load quantification can greatly improve access to viral monitoring for HIV-infected patients receiving treatment in resource-limited settings. OBJECTIVES: To evaluate and validate HIV viral load measurement from DBS in sub-Saharan Africa, with a reliable, all-automated, standard commercial assay such as the Abbott m2000. METHODS: A total of 277 DBS were collected in different health centres in Malawi and Mozambique and analysed for viral load determination using the Abbott m2000 assay with the corresponding plasma samples as gold standard. Samples were extracted using the m2000SP automatic extractor and then processed as the plasma samples using the specific 1.0 mL HIV-RNA DBS protocol. RESULTS: Among samples with detectable HIV-RNA the correlation between viral load obtained from the paired 131 plasma and DBS samples was high (r=0.946). Overall, viral load values between DBS and plasma differed by less than 0.5 log unit in 90.1% of cases and by less than 1 log unit in 100% of cases. Using a threshold of 1 000 copies/mL (defining virological failure in resource-limited settings), sensitivity was 94.2% and specificity 98.6%, and both positive and negative predictive values were high (98.5% and 94.5%, respectively). CONCLUSION: DBS extracted and processed using the Abbott automated system can be reliably used in resource-limited setting to diagnose virological failure.

Erba, F., Brambilla, D., Ceffa, S., Ciccacci, F., Luhanga, R., Sidumo, Z., et al. (2015). Measurement of viral load by the automated Abbott real-time HIV-1 assay using dried blood spots collected and processed in Malawi and Mozambique. SAMJ SOUTH AFRICAN MEDICAL JOURNAL, 105(12), 1036-1038 [10.7196/SAMJ.2015.v105i12.9673].

Measurement of viral load by the automated Abbott real-time HIV-1 assay using dried blood spots collected and processed in Malawi and Mozambique

ERBA, FULVIO;PALOMBI, LEONARDO;MANCINELLI, SANDRO;
2015-01-01

Abstract

BACKGROUND: The use of dried blood spots (DBS) for HIV-1 viral load quantification can greatly improve access to viral monitoring for HIV-infected patients receiving treatment in resource-limited settings. OBJECTIVES: To evaluate and validate HIV viral load measurement from DBS in sub-Saharan Africa, with a reliable, all-automated, standard commercial assay such as the Abbott m2000. METHODS: A total of 277 DBS were collected in different health centres in Malawi and Mozambique and analysed for viral load determination using the Abbott m2000 assay with the corresponding plasma samples as gold standard. Samples were extracted using the m2000SP automatic extractor and then processed as the plasma samples using the specific 1.0 mL HIV-RNA DBS protocol. RESULTS: Among samples with detectable HIV-RNA the correlation between viral load obtained from the paired 131 plasma and DBS samples was high (r=0.946). Overall, viral load values between DBS and plasma differed by less than 0.5 log unit in 90.1% of cases and by less than 1 log unit in 100% of cases. Using a threshold of 1 000 copies/mL (defining virological failure in resource-limited settings), sensitivity was 94.2% and specificity 98.6%, and both positive and negative predictive values were high (98.5% and 94.5%, respectively). CONCLUSION: DBS extracted and processed using the Abbott automated system can be reliably used in resource-limited setting to diagnose virological failure.
2015
Pubblicato
Rilevanza internazionale
Articolo
Esperti anonimi
Settore MED/42 - IGIENE GENERALE E APPLICATA
Settore BIO/10 - BIOCHIMICA
English
Con Impact Factor ISI
HIV Infections/diagnosis* Malawi Mozambique Viral Load/methods*
Erba, F., Brambilla, D., Ceffa, S., Ciccacci, F., Luhanga, R., Sidumo, Z., et al. (2015). Measurement of viral load by the automated Abbott real-time HIV-1 assay using dried blood spots collected and processed in Malawi and Mozambique. SAMJ SOUTH AFRICAN MEDICAL JOURNAL, 105(12), 1036-1038 [10.7196/SAMJ.2015.v105i12.9673].
Erba, F; Brambilla, D; Ceffa, S; Ciccacci, F; Luhanga, R; Sidumo, Z; Palombi, L; Mancinelli, S; Marazzi, M; Andreotti, M; Giuliano, M
Articolo su rivista
File in questo prodotto:
File Dimensione Formato  
Measurement of viral load by the automated Abbott.pdf

accesso aperto

Licenza: Creative commons
Dimensione 162.27 kB
Formato Adobe PDF
162.27 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/172739
Citazioni
  • ???jsp.display-item.citation.pmc??? 11
  • Scopus 15
  • ???jsp.display-item.citation.isi??? 12
social impact