Aim: This study evaluated the temperature change within gutta-percha during the vertical compaction technique performed with a System-B Heat Source. Methodology: Extracted human teeth were prepared, and divided in two equal groups depending on the apex diameter (group 1: 0.20-0.25 mm and group 2: 0.30-0.35 mm). Teeth were root treated with a combination of step-back and coronal flaring instrumentation using Gates Glidden burs and hand stainless steel K- and H-files. Two thermocouples were connected to the outer surface of the root the first one was placed 2 mm apically from cement-enamel junction (CEJ) (point C); and the second one 1.5 mm from the apex (point A). The instrumented canals were filled with size F gutta-percha cones. All teeth were then immersed in a thermostatic bath at a constant temperature of 37°C and warm vertical compaction was performed using a System-B Heat Source. ΔT values were recorded by means of a digital thermometer connected to the thermocouples. Results: Increments of temperature (ΔT) recorded In point A revealed a mean value of 0.5 ± 0.5°C for group 1 and 0.9 ± 1.1°C for group 2; ΔT values recorded at point C gave a mean value of 4.1 ± 1.7°C for group 1 and 3.9 ± 1.81°C in group 2. No statistical difference was found between the groups, whilst a difference was present between the measurements at points A and C (P < 0.01). Conclusions: The use of the System-B Heat Source on root canals maintained at a constant body temperature by a thermostatic bath revealed that the increase of temperature of the gutta-percha at the apical third of the canal was negligible and that the compaction of the mass of the gutta-percha close to the apex was performed at body temperature. Minor changes in temperature of the outer surface of the root canals occurred, suggesting no danger for the periradicular tissues.

Venturi, M., Pasquantonio, G., Falconi, M., Breschi, L. (2002). Temperature change within gutta-percha induced by the System-B Heat Source. INTERNATIONAL ENDODONTIC JOURNAL, 35(9), 740-746 [10.1046/j.1365-2591.2002.00553.x].

Temperature change within gutta-percha induced by the System-B Heat Source

PASQUANTONIO, GUIDO;
2002-01-01

Abstract

Aim: This study evaluated the temperature change within gutta-percha during the vertical compaction technique performed with a System-B Heat Source. Methodology: Extracted human teeth were prepared, and divided in two equal groups depending on the apex diameter (group 1: 0.20-0.25 mm and group 2: 0.30-0.35 mm). Teeth were root treated with a combination of step-back and coronal flaring instrumentation using Gates Glidden burs and hand stainless steel K- and H-files. Two thermocouples were connected to the outer surface of the root the first one was placed 2 mm apically from cement-enamel junction (CEJ) (point C); and the second one 1.5 mm from the apex (point A). The instrumented canals were filled with size F gutta-percha cones. All teeth were then immersed in a thermostatic bath at a constant temperature of 37°C and warm vertical compaction was performed using a System-B Heat Source. ΔT values were recorded by means of a digital thermometer connected to the thermocouples. Results: Increments of temperature (ΔT) recorded In point A revealed a mean value of 0.5 ± 0.5°C for group 1 and 0.9 ± 1.1°C for group 2; ΔT values recorded at point C gave a mean value of 4.1 ± 1.7°C for group 1 and 3.9 ± 1.81°C in group 2. No statistical difference was found between the groups, whilst a difference was present between the measurements at points A and C (P < 0.01). Conclusions: The use of the System-B Heat Source on root canals maintained at a constant body temperature by a thermostatic bath revealed that the increase of temperature of the gutta-percha at the apical third of the canal was negligible and that the compaction of the mass of the gutta-percha close to the apex was performed at body temperature. Minor changes in temperature of the outer surface of the root canals occurred, suggesting no danger for the periradicular tissues.
2002
Pubblicato
Rilevanza internazionale
Articolo
Esperti anonimi
Settore MED/28 - MALATTIE ODONTOSTOMATOLOGICHE
English
Con Impact Factor ISI
Gutta-percha; Root canal; System-B heat source; Temperature; Dentistry (all)
Venturi, M., Pasquantonio, G., Falconi, M., Breschi, L. (2002). Temperature change within gutta-percha induced by the System-B Heat Source. INTERNATIONAL ENDODONTIC JOURNAL, 35(9), 740-746 [10.1046/j.1365-2591.2002.00553.x].
Venturi, M; Pasquantonio, G; Falconi, M; Breschi, L
Articolo su rivista
File in questo prodotto:
File Dimensione Formato  
Int_Endod_J_2002_Venturi.pdf

solo utenti autorizzati

Licenza: Copyright dell'editore
Dimensione 137.21 kB
Formato Adobe PDF
137.21 kB Adobe PDF   Visualizza/Apri   Richiedi una copia

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/169188
Citazioni
  • ???jsp.display-item.citation.pmc??? 9
  • Scopus 52
  • ???jsp.display-item.citation.isi??? 49
social impact