Purpose: The aim of the present work was to evaluate small field size output factors (OFs) using the latest diamond detector commercially available, PTW-60019 microDiamond, over different CyberKnife systems. OFs were measured also by silicon detectors routinely used by each center, considered as reference. Methods: Five Italian CyberKnife centers performed OFs measurements for field sizes ranging from 5 to 60 mm, defined by fixed circular collimators (5 centers) and by Iris™ variable aperture collimator (4 centers). Setup conditions were: 80 cm source to detector distance, and 1.5 cm depth in water. To speed up measurements two diamond detectors were used and their equivalence was evaluated. MonteCarlo (MC) correction factors for silicon detectors were used for comparing the OF measurements. Results: Considering OFs values averaged over all centers, diamond data resulted lower than uncorrected silicon diode ones. The agreement between diamond and MC corrected silicon values was within 0.6% for all fixed circular collimators. Relative differences between microDiamond and MC corrected silicon diodes data for Iris™ collimator were lower than 1.0% for all apertures in the totality of centers. The two microDiamond detectors showed similar characteristics, in agreement with the technical specifications. Conclusions: Excellent agreement between microDiamond and MC corrected silicon diode detectors OFs was obtained for both collimation systems fixed cones and Iris™, demonstrating the microDiamond could be a suitable detector for CyberKnife commissioning and routine checks. These results obtained in five centers suggest that for CyberKnife systems microDiamond can be used without corrections even at the smallest field size. © 2016 Associazione Italiana di Fisica Medica.

Russo, S., Masi, L., Francescon, P., Frassanito, M., Fumagalli, M., Marinelli, M., et al. (2016). Multicenter evaluation of a synthetic single-crystal diamond detector for CyberKnife small field size output factors. PHYSICA MEDICA, 32(4), 575-581 [10.1016/j.ejmp.2016.03.005].

Multicenter evaluation of a synthetic single-crystal diamond detector for CyberKnife small field size output factors

MARINELLI, MARCO;VERONA RINATI, GIANLUCA;
2016-01-01

Abstract

Purpose: The aim of the present work was to evaluate small field size output factors (OFs) using the latest diamond detector commercially available, PTW-60019 microDiamond, over different CyberKnife systems. OFs were measured also by silicon detectors routinely used by each center, considered as reference. Methods: Five Italian CyberKnife centers performed OFs measurements for field sizes ranging from 5 to 60 mm, defined by fixed circular collimators (5 centers) and by Iris™ variable aperture collimator (4 centers). Setup conditions were: 80 cm source to detector distance, and 1.5 cm depth in water. To speed up measurements two diamond detectors were used and their equivalence was evaluated. MonteCarlo (MC) correction factors for silicon detectors were used for comparing the OF measurements. Results: Considering OFs values averaged over all centers, diamond data resulted lower than uncorrected silicon diode ones. The agreement between diamond and MC corrected silicon values was within 0.6% for all fixed circular collimators. Relative differences between microDiamond and MC corrected silicon diodes data for Iris™ collimator were lower than 1.0% for all apertures in the totality of centers. The two microDiamond detectors showed similar characteristics, in agreement with the technical specifications. Conclusions: Excellent agreement between microDiamond and MC corrected silicon diode detectors OFs was obtained for both collimation systems fixed cones and Iris™, demonstrating the microDiamond could be a suitable detector for CyberKnife commissioning and routine checks. These results obtained in five centers suggest that for CyberKnife systems microDiamond can be used without corrections even at the smallest field size. © 2016 Associazione Italiana di Fisica Medica.
2016
Pubblicato
Rilevanza internazionale
Articolo
Esperti anonimi
Settore FIS/07 - FISICA APPLICATA (A BENI CULTURALI, AMBIENTALI, BIOLOGIA E MEDICINA)
English
CyberKnife; Multicenter evaluation; Output factor; Single microDiamond detector; Small beam dosimetry
https://www.scopus.com/inward/record.uri?eid=2-s2.0-84961962680&partnerID=40&md5=059eae65f33df2d3ae31c886e2b7ec3d
Russo, S., Masi, L., Francescon, P., Frassanito, M., Fumagalli, M., Marinelli, M., et al. (2016). Multicenter evaluation of a synthetic single-crystal diamond detector for CyberKnife small field size output factors. PHYSICA MEDICA, 32(4), 575-581 [10.1016/j.ejmp.2016.03.005].
Russo, S; Masi, L; Francescon, P; Frassanito, M; Fumagalli, M; Marinelli, M; Falco, M; Martinotti, A; Pimpinella, M; Reggiori, G; VERONA RINATI, G; Vi...espandi
Articolo su rivista
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/166657
Citazioni
  • ???jsp.display-item.citation.pmc??? ND
  • Scopus 31
  • ???jsp.display-item.citation.isi??? 31
social impact