This paper describes a new design of microfluidic impedance cytometer enabling accurate characterization of particles without the need for focusing. The approach uses multiple pairs of electrodes to measure the transit time of particles through the device in two simultaneous different current measurements, a transverse (top to bottom) current and an oblique current. This gives a new metric that can be used to estimate the vertical position of the particle trajectory through the microchannel. This parameter effectively compensates for the non-uniform electric field in the channel that is an unavoidable consequence of the use of planar parallel facing electrodes. The new technique is explained and validated using numerical modelling. Impedance data for 5, 6 and 7 μm particles are collected and compared with simulations. The method gives excellent coefficient of variation in (electrical) radius of particles of 1% for a sheathless configuration.

Spencer, D., Caselli, F., Bisegna, P., Morgan, H. (2016). High accuracy particle analysis using sheathless microfluidic impedance cytometry. LAB ON A CHIP, 16(13), 2467-2473 [10.1039/c6lc00339g].

High accuracy particle analysis using sheathless microfluidic impedance cytometry

CASELLI, FEDERICA;BISEGNA, PAOLO;
2016-05-31

Abstract

This paper describes a new design of microfluidic impedance cytometer enabling accurate characterization of particles without the need for focusing. The approach uses multiple pairs of electrodes to measure the transit time of particles through the device in two simultaneous different current measurements, a transverse (top to bottom) current and an oblique current. This gives a new metric that can be used to estimate the vertical position of the particle trajectory through the microchannel. This parameter effectively compensates for the non-uniform electric field in the channel that is an unavoidable consequence of the use of planar parallel facing electrodes. The new technique is explained and validated using numerical modelling. Impedance data for 5, 6 and 7 μm particles are collected and compared with simulations. The method gives excellent coefficient of variation in (electrical) radius of particles of 1% for a sheathless configuration.
31-mag-2016
Pubblicato
Rilevanza internazionale
Articolo
Esperti anonimi
Settore ING-IND/34 - BIOINGEGNERIA INDUSTRIALE
English
Con Impact Factor ISI
microfluidic impedance cytometry; single-cell analysis; size estimate; signal processing; modeling and simulation; chip design
Accesso Aperto MIUR. The data from this paper can be obtained from the University of Southampton repository at http://dx.doi.org/10.5258/SOTON/395307.
Spencer, D., Caselli, F., Bisegna, P., Morgan, H. (2016). High accuracy particle analysis using sheathless microfluidic impedance cytometry. LAB ON A CHIP, 16(13), 2467-2473 [10.1039/c6lc00339g].
Spencer, D; Caselli, F; Bisegna, P; Morgan, H
Articolo su rivista
File in questo prodotto:
File Dimensione Formato  
High accuracy particle analysis.pdf

Open Access dal 01/06/2017

Descrizione: Articolo principale
Licenza: Copyright dell'editore
Dimensione 13.14 MB
Formato Adobe PDF
13.14 MB Adobe PDF Visualizza/Apri