Purpose: To investigate for the first time the dosimetric properties of a new commercial synthetic diamond detector (PTW microDiamond) in high-energy scanned clinical carbon ion beams generated by a synchrotron at the CNAO facility. Methods: The detector response was evaluated in a water phantom with actively scanned carbon ion beams ranging from 115 to 380 MeV/u (30-250 mm Bragg peak depth in water). Homogeneous square fields of 3×3 and 6×6 cm2 were used. Short- and medium-term (2 months) detector response stability, dependence on beam energy as well as ion type (carbon ions and protons), linearity with dose, and directional and dose-rate dependence were investigated. The depth dose curve of a 280 MeV/u carbon ion beam, scanned over a 3×3 cm<sup>2</sup> area, was measured with the microDiamond detector and compared to that measured using a PTW Advanced Markus ionization chamber, and also simulated using FLUKA Monte Carlo code. The detector response in two spread-out-Bragg-peaks (SOBPs), respectively, centered at 9 and 21 cm depths in water and calculated using the treatment planning system (TPS) used at CNAO, was measured. Results: A negligible drift of detector sensitivity within the experimental session was seen, indicating that no detector preirradiation was needed. Short-term response reproducibility around 1% (1 standard deviation) was found. Only 2% maximum variation of microDiamond sensitivity was observed among all the evaluated proton and carbon ion beam energies. The detector response showed a good linear behavior. Detector sensitivity was found to be dose-rate independent, with a variation below 1.3% in the evaluated dose-rate range. A very good agreement between measured and simulated Bragg curves with both microDiamond and Advanced Markus chamber was found, showing a negligible LET dependence of the tested detector. A depth dose curve was also measured by positioning the microDiamond with its main axis oriented orthogonally to the beam direction. A strong distortion in Bragg peak measurement was observed, confirming manufacturer recommendation on avoiding such configuration. Very good results were obtained for SOBP measurements, with a difference below 1% between measured and TPS-calculated doses. The stability of detector sensitivity in the observation period was within the experimental uncertainty. Conclusions: Dosimetric characterization of a PTW microDiamond detector in high-energy scanned carbon ion beams was performed. The results of the present study showed that this detector is suitable for dosimetry of clinical carbon ion beams, with a negligible LET and dose-rate dependence.

Marinelli, M., Prestopino, G., Verona, C., VERONA RINATI, G., Ciocca, M., Mirandola, A., et al. (2015). Dosimetric characterization of a microDiamond detector in clinical scanned carbon ion beams. MEDICAL PHYSICS, 42(4), 2085-2093 [10.1118/1.4915544].

Dosimetric characterization of a microDiamond detector in clinical scanned carbon ion beams

MARINELLI, MARCO;PRESTOPINO, GIUSEPPE;VERONA, CLAUDIO;VERONA RINATI, GIANLUCA;
2015-01-01

Abstract

Purpose: To investigate for the first time the dosimetric properties of a new commercial synthetic diamond detector (PTW microDiamond) in high-energy scanned clinical carbon ion beams generated by a synchrotron at the CNAO facility. Methods: The detector response was evaluated in a water phantom with actively scanned carbon ion beams ranging from 115 to 380 MeV/u (30-250 mm Bragg peak depth in water). Homogeneous square fields of 3×3 and 6×6 cm2 were used. Short- and medium-term (2 months) detector response stability, dependence on beam energy as well as ion type (carbon ions and protons), linearity with dose, and directional and dose-rate dependence were investigated. The depth dose curve of a 280 MeV/u carbon ion beam, scanned over a 3×3 cm2 area, was measured with the microDiamond detector and compared to that measured using a PTW Advanced Markus ionization chamber, and also simulated using FLUKA Monte Carlo code. The detector response in two spread-out-Bragg-peaks (SOBPs), respectively, centered at 9 and 21 cm depths in water and calculated using the treatment planning system (TPS) used at CNAO, was measured. Results: A negligible drift of detector sensitivity within the experimental session was seen, indicating that no detector preirradiation was needed. Short-term response reproducibility around 1% (1 standard deviation) was found. Only 2% maximum variation of microDiamond sensitivity was observed among all the evaluated proton and carbon ion beam energies. The detector response showed a good linear behavior. Detector sensitivity was found to be dose-rate independent, with a variation below 1.3% in the evaluated dose-rate range. A very good agreement between measured and simulated Bragg curves with both microDiamond and Advanced Markus chamber was found, showing a negligible LET dependence of the tested detector. A depth dose curve was also measured by positioning the microDiamond with its main axis oriented orthogonally to the beam direction. A strong distortion in Bragg peak measurement was observed, confirming manufacturer recommendation on avoiding such configuration. Very good results were obtained for SOBP measurements, with a difference below 1% between measured and TPS-calculated doses. The stability of detector sensitivity in the observation period was within the experimental uncertainty. Conclusions: Dosimetric characterization of a PTW microDiamond detector in high-energy scanned carbon ion beams was performed. The results of the present study showed that this detector is suitable for dosimetry of clinical carbon ion beams, with a negligible LET and dose-rate dependence.
2015
Pubblicato
Rilevanza internazionale
Articolo
Esperti anonimi
Settore FIS/07 - FISICA APPLICATA (A BENI CULTURALI, AMBIENTALI, BIOLOGIA E MEDICINA)
English
carbon ions; relative dosimetry; synthetic diamond detector; Biophysics; Radiology, Nuclear Medicine and Imaging
http://scitation.aip.org/content/aapm/journal/medphys
Marinelli, M., Prestopino, G., Verona, C., VERONA RINATI, G., Ciocca, M., Mirandola, A., et al. (2015). Dosimetric characterization of a microDiamond detector in clinical scanned carbon ion beams. MEDICAL PHYSICS, 42(4), 2085-2093 [10.1118/1.4915544].
Marinelli, M; Prestopino, G; Verona, C; VERONA RINATI, G; Ciocca, M; Mirandola, A; Mairani, A; Raffaele, L; Magro, G
Articolo su rivista
File in questo prodotto:
File Dimensione Formato  
146.pdf

accesso aperto

Licenza: Copyright dell'editore
Dimensione 404.34 kB
Formato Adobe PDF
404.34 kB Adobe PDF Visualizza/Apri

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/113480
Citazioni
  • ???jsp.display-item.citation.pmc??? 0
  • Scopus 31
  • ???jsp.display-item.citation.isi??? 28
social impact