Purpose Aim of this work is the assessment of build-up and superficial doses of different clinical Head&Neck plans delivered with Helical TomoTherapy (HT) (Accuray, Sunnyvale, CA). Depth dose profiles and superficial dose points were measured in order to evaluate the Treatment Planning System (TPS) capability of an accurate dose modeling in regions of disequilibrium. Geometries and scattering conditions were investigated, similar to the ones generally encountered in clinical treatments. Methods Measurements were performed with two dosimeters: Gafchromic® EBT3 films (Ashland Inc., Wayne, NJ) and a synthetic single crystal diamond detector (PTW-Frieburg microDiamond, MD). A modified version of the Alderson RANDO phantom was employed to house the detectors. A comparison with TPS data was carried out in terms of dose difference (DD) and distance-to-agreement (DTA). Results DD between calculated data and MD measurements are within 4% even in points with high spatial dose variation. For depth profiles, EBT3 data show a DDmax of 3.3% and DTAmax of 2.2 mm, in low and high gradient regions, respectively, and compare well with MD data. EBT3 superficial points always results in measured doses lower than TPS evaluated ones, with a maximum DTA value of 1.5 mm. Conclusions Doses measured with the two devices are in good agreement and compare well with calculated data. The deviations found in the present work are within the reference tolerance level, suggesting that the HT TPS is capable of a precise dose estimation both in superficial regions and in correspondence with interfaces between air and PMMA.

Zani, M., Talamonti, C., Bucciolini, M., Marinelli, M., VERONA RINATI, G., Bonomo, P., et al. (2016). In phantom assessment of superficial doses under TomoTherapy irradiation. PHYSICA MEDICA, 32(10), 1263-1270 [10.1016/j.ejmp.2016.09.017].

In phantom assessment of superficial doses under TomoTherapy irradiation

MARINELLI, MARCO;VERONA RINATI, GIANLUCA;
2016-01-01

Abstract

Purpose Aim of this work is the assessment of build-up and superficial doses of different clinical Head&Neck plans delivered with Helical TomoTherapy (HT) (Accuray, Sunnyvale, CA). Depth dose profiles and superficial dose points were measured in order to evaluate the Treatment Planning System (TPS) capability of an accurate dose modeling in regions of disequilibrium. Geometries and scattering conditions were investigated, similar to the ones generally encountered in clinical treatments. Methods Measurements were performed with two dosimeters: Gafchromic® EBT3 films (Ashland Inc., Wayne, NJ) and a synthetic single crystal diamond detector (PTW-Frieburg microDiamond, MD). A modified version of the Alderson RANDO phantom was employed to house the detectors. A comparison with TPS data was carried out in terms of dose difference (DD) and distance-to-agreement (DTA). Results DD between calculated data and MD measurements are within 4% even in points with high spatial dose variation. For depth profiles, EBT3 data show a DDmax of 3.3% and DTAmax of 2.2 mm, in low and high gradient regions, respectively, and compare well with MD data. EBT3 superficial points always results in measured doses lower than TPS evaluated ones, with a maximum DTA value of 1.5 mm. Conclusions Doses measured with the two devices are in good agreement and compare well with calculated data. The deviations found in the present work are within the reference tolerance level, suggesting that the HT TPS is capable of a precise dose estimation both in superficial regions and in correspondence with interfaces between air and PMMA.
2016
Pubblicato
Rilevanza internazionale
Articolo
Esperti anonimi
Settore FIS/07 - FISICA APPLICATA (A BENI CULTURALI, AMBIENTALI, BIOLOGIA E MEDICINA)
English
Diamond detector; Radiochromic films; Superficial dosimetry; TomoTherapy; Biophysics; Radiology, Nuclear Medicine and Imaging; Physics and Astronomy (all)
www.fisicamedica.org
Zani, M., Talamonti, C., Bucciolini, M., Marinelli, M., VERONA RINATI, G., Bonomo, P., et al. (2016). In phantom assessment of superficial doses under TomoTherapy irradiation. PHYSICA MEDICA, 32(10), 1263-1270 [10.1016/j.ejmp.2016.09.017].
Zani, M; Talamonti, C; Bucciolini, M; Marinelli, M; VERONA RINATI, G; Bonomo, P; Paiar, F; Pallotta, S
Articolo su rivista
File in questo prodotto:
Non ci sono file associati a questo prodotto.

I documenti in IRIS sono protetti da copyright e tutti i diritti sono riservati, salvo diversa indicazione.

Utilizza questo identificativo per citare o creare un link a questo documento: https://hdl.handle.net/2108/168990
Citazioni
  • ???jsp.display-item.citation.pmc??? 2
  • Scopus 4
  • ???jsp.display-item.citation.isi??? 4
social impact